It has been shown that radiochromic film is an ideal dosimeter for assessment and verification of delivered dose to irradiated blood products. Using a parallel opposing two-field technique on a medical linear accelerator, blood is irradiated to diminish the risk of transfusion-associated graft vs. host disease (TA-GVHD). The blood products are irradiated in a Perspex blood box to an applied dos...
PURPOSE The aim of this study was to combine a set of methods for use of radiochromic film dosimetry, including calibration, correction for lateral effects and a proposed triple-channel analysis. These methods can be applied to GAFCHROMIC EBT3 film dosimetry for radiation field analysis and verification of IMRT plans. METHODS A single-film exposure was used to achieve dose calibration, and th...
Patient-specific verification of intensity-modulated radiation therapy (IMRT) plans can be done by dosimetric measurements or by independent dose or monitor unit calculations. The aim of this study was the clinical evaluation of IMRT verification based on a fast Monte Carlo (MC) program with regard to possible benefits compared to commonly used film dosimetry. 25 head-and-neck IMRT plans were r...
An anthropomorphic head phantom, constructed from a water-equivalent plastic shell with only a spherical target, was modified to include a nonspherical target (pituitary) and an adjacent organ at risk (OAR) (optic chiasm), within 2 mm, simulating the anatomy encountered when treating acromegaly. The target and OAR spatial proximity provided a more realistic treatment planning and dose delivery ...
Heterogeneities existing in the patient during treatment are neglected, as the treated subject is considered homogeneous in most of the commercially-available treatment planning systems (TPSs) used for high dose rate (HDR) brachytherapy. The choice of a suitable dosimeter for experimental dosimetry near the HDR source is crucial, mainly due to existence of steep dose gradients. The present work...
The purpose of this study was to evaluate a single-scan protocol using Gafchromic EBT3 film (EBT3) by comparing it with the commonly used 24-hr measurement protocol for radiochromic film dosimetry. Radiochromic film is generally scanned 24 hr after film exposure (24-hr protocol). The single-scan protocol enables measurement results within a short time using only the verification film, one calib...
Introduction: Nowadays radiotherapy plays an important role in cancer treatment. Different radiotherapy techniques improvement emphasizes on using of the precise ، appropriate and useful algorithms. one of these techniques are wedged which is used in radiotherapy to compensate missing tissues and create a uniform dose distribution in tissues. The Siemens Artiste linear accelera...
abstractbackgrounds: md-55-2 is one of the radiochromic film models with the sensitivity suitable for dose measurements ranging from 5 to 100 gy. however, this lower limit makes the film impractical for its applications in many areas such as brachytherapy source dosimetry.materials and methods: in this project, the useful range of the film has been extended by using a multilayer film technique...
Background: Accurate small radiation field dosimetry is essential in modern radiotherapy techniques such as stereotactic radiosurgery (SRS) and intensity modulated radiotherapy (IMRT). Precise measurement of dosimetric parameters such as beam profile, percentage depth doses and output factor of these beams are complicated due to the electron disequilibrium and the steep dose gradients. In the p...