Magnetic resonance (MR) imaging of Gafchromic film causes perturbation to absolute dosimetry measurements; the purpose of this work was to characterize the perturbation and develop a correction method for it. Three sets of Gafchromic EBT2 film were compared: radiation (control), radiation followed by MR imaging (RAD + B), and MR imaging followed by radiation (B + RAD). The T1-weighted and T2-we...

Journal: :Medical physics 2012
N Gopishankar R K Bisht

PURPOSE To present EBT2 film verification of treatment planning with the eXtend System, a relocatable frame system for multiple-fraction or serial multiple-session radiosurgery. METHODS A human head shaped phantom simulated the verification process for fractionated Gamma Knife (GK) treatment. Phantom preparation for eXtend Frame based treatment planning involved creating a dental impression, ...

2016
Farideh Farokhi Moghadam Hozan Mohammadi

Small photon field used in the intensity modulated radiotherapy (IMRT) and stereotactic radio surgery (SRS). It is difficult to measure beam profile and output factor because of electron disequilibrium, partial shutting of source because of this challenges, we need the good detector. This article is collection of 5 or more article and summary of the results of EBT3 that compared with EBT2 film ...

2012
Frédéric Girard Hugo Bouchard Frédéric Lacroix

The objectives of this study are to identify and quantify factors that influence radiochromic film dose response and to determine whether such films are suitable for reference dosimetry. The influence of several parameters that may introduce systematic dose errors when performing reference dose measurements were investigated. The effect of the film storage temperature was determined by comparin...

Journal: :Medical physics 2012
N Corradini S Presilla M Sassowsky M Schombourg

PURPOSE To establish a film dosimetric method for high-resolution measurement in the dose build-up region. METHODS Percent depth dose (PDD) curves were measured in water for TomoTherapy using Gafchromic EBT2 films. Depth dose measurements were also performed using Standard Imaging A1SL and PTW PinPoint ionization chambers, as well as a PTW DiodeE detector. The film and detector measurements w...

2012
Hirokazu Mizuno Yutaka Takahashi Atsushi Tanaka Takamitsu Hirayama Tsuyoshi Yamaguchi Hiroaki Katou Keiko Takahara Yoshiaki Okamoto Teruki Teshima

EBT2 film is widely used for quality assurance in radiation therapy. The purpose of this study was to investigate the homogeneity of EBT2 film among various lots, and the dose dependence of heterogeneity. EBT2 film was positioned in the center of a flatbed scanner and scanned in transmission mode at 75 dpi. Homogeneity was investigated by evaluating gray value and net optical density (netOD) wi...

2012
Naoki Hayashi Yoichi Watanabe Ryan Malmin Hideki Kato

The purpose of this study was to evaluate the triple channel correction acquisition (TCCA) method for radiochromic film dosimetry performed with a flatbed scanner. The study had two parts: a fundamental and a clinical examination. In the fundamental examination, we evaluated the accuracy of calibration curves for Gafchromic EBT2 (EBT2). The films were calibrated using a field-by-field method wi...

Journal: :Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics 2014
M Zarza-Moreno P Carreira L Madureira H Miras Del Rio F J Salguero A Leal N Teixeira A P Jesus G Mora

This study evaluates the dosimetric impact caused by an air cavity located at 2 mm depth from the top surface in a PMMA phantom irradiated by electron beams produced by a Siemens Primus linear accelerator. A systematic evaluation of the effect related to the cavity area and thickness as well as to the electron beam energy was performed by using Monte Carlo simulations (EGSnrc code), Pencil Beam...