نتایج جستجو برای: hotelling observer

تعداد نتایج: 28725  

2016
Julien G. Ott Alexandre Ba Damien Racine Nick Ryckx François O. Bochud Hatem Alkadhi Francis R. Verdun

The goal of the present work was to report and investigate the performances of a new iterative reconstruction algorithm, using a model observer. For that, a dedicated low-contrast phantom containing different targets was scanned at four volume computed tomography dose index (CTDIvol) levels on a Siemens SOMATOM Force computed tomography (CT). The acquired images were reconstructed using the ADM...

Journal: :Physics in medicine and biology 2006
I-Y Son M Winslow B Yazici X G Xu

This study develops and demonstrates a realistic x-ray imaging simulator with computerized observers to maximize lesion detectability and minimize patient exposure. A software package, ViPRIS, incorporating two computational patient phantoms, has been developed for simulating x-ray radiographic images. A tomographic phantom, VIP-Man, constructed from Visible Human anatomical colour images is us...

Journal: :Journal of the Optical Society of America. A, Optics, image science, and vision 2014
Di Ge Lu Zhang Christine Cavaro-Ménard Patrick Le Callet

This paper addresses the numerical stability issue on the channelized Hotelling observer (CHO). The CHO is a well-known approach in the medical image quality assessment domain. Many researchers have found that the detection performance of the CHO does not increase with the number of channels, contrary to expectation. And to our knowledge, nobody in this domain has found the reason. We illustrat...

Journal: :Journal of medical imaging 2023

PurposeThe objective assessment of image quality (IQ) has been advocated for the analysis and optimization medical imaging systems. One method computing such IQ metrics is through a numerical observer. The Hotelling observer (HO) optimal linear observer, but conventional methods obtaining HO can become intractable due to large sizes or insufficient data. Channelized are sometimes employed in ci...

2011
Mahdi M. Kalayeh Thibault Marin P. Hendrik Pretorius Miles N. Wernick Yongyi Yang Jovan G. Brankov

In this paper, we present a numerical observer for image quality assessment, aiming to predict human observer accuracy in a cardiac perfusion defect detection task for single-photon emission computed tomography (SPECT). In medical imaging, image quality should be assessed by evaluating the human observer accuracy for a specific diagnostic task. This approach is known as task-based assessment. S...

نمودار تعداد نتایج جستجو در هر سال

با کلیک روی نمودار نتایج را به سال انتشار فیلتر کنید