نتایج جستجو برای: modulated
تعداد نتایج: 60877 فیلتر نتایج به سال:
BACKGROUND The multileaf collimator (MLC) is a critical component to accurate intensity-modulated radiotherapy (IMRT) delivery. This study examined MLC positional accuracy via MLC logs from Varian machines from six institutions and three delivery techniques to evaluate typical positional accuracy and treatment and mechanical parameters that affect accuracy. Typical accuracy achieved was compare...
A patient specific quality assurance program has been developed to facilitate the clinical implementation of intensity modulated radiotherapy (IMRT) delivered using a micro-multileaf collimator. The methodology includes several dosimetric tasks that are performed prior to the treatment of each patient. Film dosimetry is performed for each individual field and for the multifield composite plan. ...
One of the applications of MatriXX (IBA Dosimetry) is experimental verification of dose for IMRT, VMAT, and tomotherapy. For cumulative plan verification, dose is delivered for all the treatment gantry angles to a stationary detector. Experimental calibration of MatriXX detector recommended by the manufacturer involves only AP calibration fields and does not address angular dependency of MatriX...
Volumetric-modulated arc therapy (VMAT) plans may require more control points (or segments) than some of fixed-beam IMRT plans that are created with a limited number of segments. Increasing number of control points in a VMAT plan for a given prescription dose could create a large portion of the total number of segments with small number monitor units (MUs) per segment. The purpose of this study...
BACKGROUND An advantage of the Intensity Modulated Radiotherapy (IMRT) technique is the feasibility to deliver different therapeutic dose levels to PTVs in a single treatment session using the Simultaneous Integrated Boost (SIB) technique. The paper aims to describe an automated tool to calculate the dose to be delivered with the SIB-IMRT technique in different anatomical regions that have the ...
This work applied statistical process control to establish the control limits of the % gamma pass of patient-specific intensity modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) quality assurance (QA), and to evaluate the efficiency of the QA process by using the process capability index (Cpml). A total of 278 IMRT QA plans in nasopharyngeal carcinoma were measured with ...
PURPOSE The purpose of this review is to address the issue of unnecessary irradiation of the normal larynx. METHODS We performed a literature review and dosimetry evaluation of intensity modulated radiotheraphy (IMRT) techniques. RESULTS Data suggest that matching conventional parallel opposed photon fields through adenopathy is a reasonable alternative to the common practice of unnecessari...
The purpose of this study was to measure the multileaf collimator (MLC) transmission from the first Hi-Art II tomotherapy machine installed at the Advanced Center for Treatment, Research, and Education in Cancer (ACTREC). The MLC transmission was measured with an A1SL ion chamber and the radiographic extended dose range (EDR2) film in virtual water slabs at 1.5-cm depth with a source-to-surface...
PURPOSE Several authors have recommended a 2 mm tolerance for multileaf collimator (MLC) positioning in sliding window treatments. In volumetric modulated arc therapy (VMAT) treatments, however, the optimal tolerance for MLC positioning remains unknown. In this paper, the authors present the results of a multicenter study to determine the optimal tolerance for both techniques. METHODS The pro...