This study evaluated the agreement of fiducial marker localization between two modalities--an electronic portal imaging device (EPID) and cone-beam computed tomography (CBCT)--using a low-dose, half-rotation scanning protocol. Twenty-five prostate cancer patients with implanted fiducial markers were enrolled. Before each daily treatment, EPID and half-rotation CBCT images were acquired. Transla...
PURPOSE The aim of this study is to investigate the performance and limitations of a real-time transit electronic portal imaging device (EPID) dosimetry system for error detection during dynamic intensity modulated radiation therapy (IMRT) treatment delivery. Sites studied are prostate, head and neck (HN), and rectal cancer treatments. METHODS The system compares measured cumulative transit E...
In-vivo dosimetry techniques are currently being applied only by a few Centers because they require time-consuming implementation measurements, and workload for detector positioning and data analysis. The transit in-vivo dosimetry performed by the electronic portal imaging device (EPID) avoids the problem of solid-state detector positioning on the patient. Moreover, the dosimetric characterizat...
The Electronic Portal Imaging Device (EPID) provides localization quality images and computer-aided analysis, which should in principal, replace portal film imaging. Modern EPIDs deliver superior image quality and an array of analysis tools that improve clinical decision making. It has been demonstrated that the EPID can be a powerful tool in the reduction of treatment setup errors and the qual...
Purpose: Kilovoltage cone-beam computed tomography (CBCT) has been developed to provide accurate soft-tissue and bony setup information. We evaluated clinical CBCT setup data and compared CBCT measurements with electronic portal imaging device (EPID) images. Material and Methods: The setup error for CBCT scans at the treatment unit relative to the planning CT was measured for 62 patients (524 s...
Electronic Portal Imaging Devices (EPID) are used for the assessment of the geometric accuracy 2 during the radiation therapy process. EPID is also used like a major detector to test performances of the linear accelerator used for the treatment. Therefore, the quality control of portal images 4, 5, 6, 7 is essential for this task. Software tools that are capable to perform an automatic quality ...
The feasibility of using portal dosimetry (PD) to verify 6 MV flattening filter-free (FFF) IMRT treatments was investigated. An Elekta Synergy linear accelerator with an Agility collimator capable of delivering FFF beams and a standard iViewGT amorphous silicon (aSi) EPID panel (RID 1640 AL5P) at a fixed SSD of 160 cm were used. Dose rates for FFF beams are up to four times higher than for conv...
PURPOSE VMAT involves multi-parameter modulations, thus presenting multi-dimensional challenges. Here, we present a viable approach to VMAT patient specific QA using EPID-based dosimetry for spine VMAT SRS casesMethods: Our spine VMAT SRS plans consisted of two full arcs and were optimized with 6 MV photons for a Varian Trilogy LINAC. The QA plans were computed by resetting all control points t...
PURPOSE To investigate various issues for clinical implementation of aSi EPID panels for IMRT/VMAT QA. METHODS Six linacs are used in our clinic for EPID-based plan QA; two Varian Truebeams, two Varian 2100 series, two Elekta Infiniti series. Multiple corrections must be accounted for in the calibration of each panel for dosimetric use. Varian aSi panels are calibrated with standard dark fiel...
Varian introduced (in 2010) the option of removing the flattening filter (FF) in their C-Arm linacs for intensity-modulated treatments. This mode, called flattening filter-free (FFF), offers the advantage of a greater dose rate. Varian's "Portal Dosimetry" is an electronic portal imager device (EPID)-based tool for IMRT verification. This tool lacks the capability of verifying flattening filter...