A procedure for interactive navigation of a fully motorized surgical microscope is presented. This procedure uses preoperative MRTor CT-images to generate a 3D-model of the patient which then acts as the basis for navigation during surgery. The relation between tracking system, microscope and 3D-model is established using a simple calibration method based on the Horn [Ho87] algorithm. It enable...