A-142 Comparison of Alinity c Hemoglobin A1c Immunoassay with Premier Hb9210 Automated HPLC Assay: Very Few Patients May be Outside Reportable Range Using Alinity c.
نویسندگان
چکیده
Abstract Background Hemoglobin A1c (HbA1c; glycosylated hemoglobin) is used for diagnosis and monitoring of patients with diabetes. In the clinical laboratory both immunoassays high-performance liquid chromatography (HPLC) based tests are available such measurement. We use Premier Hb9210 analyzer (HPLC method; Trinity Biotech, Jamestown, NY) measuring HBA1c in whole blood. As our transitioning to Abbott chemistry systems, we compared HbA1c values obtained by Alinity c Hb9210. Methods The on boronate affinity high performance output 66 sec. analytical measurement range 3.8 18.5% although as low 2% can be reported but result may not accurate. assay measures total hemoglobin (enzymatic assay) blood then calculates %HbA1c. this 4 14% HbA1c. Specimens cannot diluted Results has excellent precision (0.4% at 5.1% 0.25% Hb 9.4%). contrast, HB9210 1.89% 5.5% 1.81% 8.4%. However, HB9219 slightly wider assay. Plotting results x-axis corresponding assay, observed following regression equation: y = 0.9499 x + 0.1108 ( n 50, 0.99). 95% confidence interval slope was 0.9241 0.9757. Rarely, get specimen where between 4.0% (42 specimens out 52 803 2/7/22 2/7/23; 0.08%) or above (174 802; same period; 0.33%) these would produce any recently 2.4 % a sickle cell patient (reported <3.8% per lab policy). Discussion Our indicates that immunoassay showed comparable HPLC method. very few exceed due uncontrollable Because diluted, real value could measured using Conclusions conclude viable alternative laboratory. (sickle trait, uncontrolled diabetes), outside c, requiring analysis.
منابع مشابه
An automated "high-pressure" liquid-chromatographic assay for hemoglobin A1c.
An automated "high-pressure" liquid-chromatographic assay for hemoglobin A1c is described. We use a 45-min incubation in acetate buffer (pH 5.5) to eliminate labile glycated hemoglobins. In this automated system conventional modules are used but it incorporates a solvent-switching valve to select either of two buffers, which differ in pH and NaCl concentration. The chromatographic column contai...
متن کاملComparison of HbA1c analysers: Agilent 1100 HPLC using kits produced by Gordion Diagnostic (Turkey) with Premier Hb9210 using kits produced by Trinity Biotech (USA) in different patient groups.
BACKGROUND The aim of this study was to compare the performance of Agilent 1100 HPLC analyser using HbA1c kits manufactured by Gordion Diagnostic (Turkey) with that of Premier Hb9210 using the original kits for the measurement of HbA1c in different patient groups. METHODS Subjects were divided into four groups: Group 1 included 140 diabetic and non-diabetic subjects with normal urea and haemo...
متن کاملHemoglobin Raleigh as the cause of a falsely increased hemoglobin A1C in an automated ion-exchange HPLC method.
Irreversible glycation of the hemoglobin A0 (HbA0) beta chain leads to the production of HbA1C, which can be used to monitor long-term blood glucose control in patients with diabetes mellitus. HbA1C is less positively charged than nonglycated HbA0, and this decrease in charge is the basis of ion-exchange and electrophoretic methods that measure HbA1C. We recently identified a sample that appear...
متن کاملFew point-of-care hemoglobin A1c assay methods meet clinical needs.
Accurate and timely measurements of hemoglobin A1c (Hb A1c) 2 are an increasingly important challenge for clinical laboratories. As Hb A1c measurements are being used to answer new clinical questions, each new use of Hb A1c imposes requirements for the analytical performance of the methods used to measure it. Until recently, Hb A1c was used only as a key test for monitoring glycemic control in ...
متن کاملPerformance of the Roche second generation hemoglobin A1c immunoassay in the presence of HB-S or HB-C traits.
Blood HbA1c determination is a powerful tool for the evaluation and management of patients with diabetes mellitus. Many HbA1c analytical methods demonstrate bias in samples from patients with hemoglobinopathies. This study evaluated the analytical performance of Roche Diagnostics' 1st and 2nd generation HbA1c assays in patients with or without hemoglobinopathies whose HbA1c levels were elevated...