Incomplete Endothelialization and Late Dislocation After Implantation of an Amplatzer Septal Occluder Device
نویسندگان
چکیده
منابع مشابه
Incomplete endothelialization and late dislocation after implantation of an Amplatzer septal occluder device.
A 64-year-old woman with a history of a large-sized secundum atrial septal defect (ASD) underwent successful percutaneous closure with a 34-mm Amplatzer septal occluder (ASO) without apparent complications on April 25, 2002. Transthoracic echocardiography immediately following the closing procedure revealed that her ASO was well wedged, with no residual shunt or murmurs (Figure 1). At the third...
متن کاملImages in Cardiovascular Medicine Incomplete Endothelialization and Late Dislocation After Implantation of an Amplatzer Septal Occluder Device
A 64-year-old woman with a history of a large-sized secundum atrial septal defect (ASD) underwent successful percutaneous closure with a 34-mm Amplatzer septal occluder (ASO) without apparent complications on April 25, 2002. Transthoracic echocardiography immediately following the closing procedure revealed that her ASO was well wedged, with no residual shunt or murmurs (Figure 1). At the third...
متن کاملLate Bacterial Endocarditis and Abscess Formation After Implantation of an Amplatzer Septal Occluder Device.
A 37-year-old man was referred to our emergency department after 4 days of a spiking fever of up to 40°C, myalgia, and new development of multiple, small, nontender erythematous macular lesions on his palms and soles compatible with Janeway lesions (Figure 1). He had a history of a large secundum atrial septal defect for which he underwent transcatheter closure with a 24-mm Amplatzer septal occ...
متن کاملImages in cardiovascular medicine. Incomplete endothelialization and late development of acute bacterial endocarditis after implantation of an Amplatzer septal occluder device.
A 4-year-old girl with a history of a moderate-sized secundum atrial septal defect and mild mitral valve prolapse underwent transcatheter closure with a 22-mm Amplatzer septal occluder (ASO) (AGA Medical Corporation, Golden Valley, Minn) device. Approximately 12 months later, she presented to the emergency department in septic shock with a 5-day history of fevers up to 105°F and was transferred...
متن کاملVery Late Dislocation of an AMPLATZER Septal Occluder Device Suspected Thanks to a Recent Onset of Right-Axis Deviation.
D ata on late complications after percutaneous closure of atrial septal defects (ASDs) are still sparse. This is the first report of a very late dislocation of an AMPLATZER septal occluder (ASO) device (St. Jude Medical, St. Paul, Minnesota) 14 years after placement. A 74-year-old Caucasian woman was admitted to our department for a syncopal event preceded by palpitations. Her medical history i...