an investigation on the performance of dose calibrators in nuclear medicine centers in iran

چکیده

background: to investigate the status of the nuclear medicine (nm) centers in iran for the performance of dose calibrators, 18 out of 54 centers providing nm services in iran were randomly selected and inspected in 1997. in the first phase of the study the selected centers were inspected for performing of quality control (qc) tests of dose calibrators. the linearity of the activity response, precision, accuracy, and the physical functions of the instruments, were studied. in the second phase of the study, carried out in 2006, 28 out of 75 nm centers were investigated for qc tests performance. materials and methods: the qc tests were performed by using standardized radio nuclides of tc-99m and cs-137 in the first phase, and tc-99m and i-131 in the second phase of the studies. standard procedures were used for carrying out the tests. results: according to the obtained results in the first phase of the study, 10 centers were found to be in unacceptable situation. following this study, all the concerned nm centers were informed about the results, and at the same time the repair and adjustment of the dose calibrators were requested. in addition, the appropriate training courses along with the qc testing manuals were provided to the centers. based on the data of the second phase of the study, only 6 nm centers were in unacceptable situation. the results indicated the effectiveness of the improvements carried out in the working procedures of the centers during interval between the two phases of investigation.

اسناد مشابه

  • An investigation on the performance of dose calibrators in nuclear medicine centers in Iran

    مقاله۲۰۰۸International Journal of Radiation ResearchH. Zamani Zeinali, N. Alirezazadeh, F. Atabi

    Background: To investigate the status of the nuclear medicine (NM) centers in Iran for the performance of dose calibrators, 18 out of 54 centers providing NM services in Iran were randomly selected and inspected in 1997. In the first phase of the study the selected centers were i

  • Estimation of the public radiation dose from ¹⁷⁷Lu in wastewater discharged by nuclear medicine centers in Tehran, Iran

    مقاله۲۰۲۶Asia Oceania Journal of Nuclear Medicine and BiologyMohammad Reza Deevband, Mohammad Reza Asgari, Mahasti Amoui, Mohammad Ali Ghodsi Rad

    Objective(s): Currently, the treatment of prostate and neuroendocrine tumors with radiopharmaceuticals derived from lutetium-177 is the focus of many radionuclide therapy centers. One of the reasons for turning to lutetium radiopharmaceuticals is their favorable dosimetry and sta

  • monitoring of iodine-125 and iodine-131 in thyroid of individuals in nuclear medicine centers of north west provinces of iran

    مقالهiranian journal of radiation researchg.h. heravi, h. garshasbi national radiation protection department, iranian nuclear regulatory authority, tehran, iran, j. karimi diba, s.k. asghari

    background: due to volatility of radioactive iodine solutions, usually internal exposure would occur in personnel that are working in nuclear medicine centers. the activities in the thyroid of individual s who work in nuclear medicine centers of north west provinces of iran were

  • Monitoring of iodine-125 and iodine-131 in thyroid of individuals in nuclear medicine centers of North West provinces of Iran

    مقاله۲۰۰۴International Journal of Radiation ResearchG.H. Heravi, H. Garshasbi, J. Karimi Diba, S.K. Asghari

    Background: Due to volatility of radioactive iodine solutions, usually internal exposure would occur in personnel that are working in nuclear medicine centers. The activities in the thyroid of individual s who work in nuclear medicine centers of North West provinces of Iran were

  • active personal dosimeter in a nuclear medicine center in yazd city, iran

    مقالهjournal of biomedical physics and engineeringf. bouzarjomehri department of medical physics, shahid sadoughi university of medical sciences, yazd, iran, v. tsapaki department of medical physics, konstantopoulio hospital, nea ionia, athens, greece

    introduction: active personal dosimeters (apds) are well accepted as useful and reliable instruments for individual dosimetry measurements. apds have many advantages compared with passive dosimeters for individual external radiation dose assessments. in routine monitoring, occupa

  • Active Personal Dosimeter in a Nuclear Medicine Center in Yazd City, Iran

    مقاله۲۰۱۱Journal of Biomedical Physics and EngineeringF. Bouzarjomehri, V. Tsapaki

    Introduction: Active personal dosimeters (APDs) are well accepted as useful and reliable instruments for individual dosimetry measurements. APDs have many advantages compared with passive dosimeters for individual external radiation dose assessments. In routine monitoring, occupa