validation of a prototype optical computed tomography system

نویسندگان

seyed salman zakariaee

mikaeil molazadeh

abbas takavar

alireza shirazi

چکیده

in radiation cancer treatments, the most of the side effects could be minimized using a proper dosimeter. gel dosimeter is the only three‑dimensional dosimeter and magnetic resonance imaging (mri) is the gold standard method for gel dosimeter readout. because of hard accessibility and high cost of sample reading by mri systems, some other alternative methods were developed. the optical computed tomography (oct) method could be considered as the most promising alternative method that has been studied widely. in the current study, gel dosimeter scanning using a prototype optical scanner and validation of this optical scanner was performed. optical absorbance of the irradiated gel samples was determined by both of conventional spectrophotometer and the fabricated oct system at 632 nm. furthermore, these irradiated vials were scanned by a 1.5 t mri. the slope of the curves was extracted as the dose‑response sensitivity. the r2‑dose sensitivity measured by mri method was 0.1904 and 0.113 for nipam and pagat gels, respectively. the optical dose sensitivity obtained by conventional spectrophotometer and the fabricated optical scanner was 0.0453 and 0.0442 for nipam gels and 0.0244 and 0.0242 for pagat gels, respectively. the scanning results of the absorbed dose values showed that the new oct and conventional spectrophotometer were in fair agreement. from the results, it could be concluded that the fabricated system is able to quantize the absorbed dose values in polymer gel samples with acceptable accuracy

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Description of a prototype emission-transmission computed tomography imaging system.

We have developed a prototype imaging system that can perform simultaneous x-ray transmission CT and SPECT phantom studies. This system employs a 23-element high-purity-germanium detector array. The detector array is coupled to a collimator with septa angled toward the focal spot of an x-ray tube. During image acquisition, the x-ray fan beam and the detector array move synchronously along an ar...

متن کامل

Mapping air contaminants indoors using a prototype computed tomography system.

Single and multiple plumes of sulphur hexafluoride were measured and mapped using a prototype open-path Fourier transform infrared spectrometer (OP-FTIR)-computed tomography (CT) system in an indoor exposure chamber. The OP-FTIR-CT system accurately mapped the position of concentration peaks at different locations in the chamber, and estimated concentrations within a range of 4-50% when compare...

متن کامل

Validation of computed tomography-based attenuation correction of deviation between theoretical and actual values for four computed tomography scanners

Objective: In this study, we aimed to validate the accuracy of computed tomography-based attenuation correction (CTAC) using the bilinear scaling method.Methods: The measured attenuation coefficient (μm) was compared to a theoretical attenuation coefficient (μt ) using four different CT scanners and an RMI 467 phantom. The effective energy of the CT beam X-rays was calculated, using the aluminu...

متن کامل

Optical coherence computed tomography

A time-resolved optical tomography, optical coherence computed tomography, is proposed to bridge the gap between diffuse optical tomography and optical coherence tomography. Both ballistic and multiple-scattered photons are measured at multiple source-detection positions by low-coherence interferometry providing a temporal resolution smaller than 100 fs. A light-tissue interaction model was est...

متن کامل

Siemens primus accelerator simulation using EGSnrc Monte Carlo code and gel dosimetry validation with optical computed tomography system by EGSnrc code

Monte Carlo method is the most accurate method for simulation of radiation therapy equipment. The linear accelerators (linac) are currently the most widely used machines in radiation therapy centers. Monte Carlo modeling of the Siemens Primus linear accelerator in 6 MeV beams was used. Square field size of 10 × 10 cm2 produced by the jaws was compared with TLD. Head simulation of Siemens accele...

متن کامل

AbstractID: 12899 Title: Dosimetric Characterization of a Dedicated Breast Computed Tomography Clinical Prototype System

ID: 12899 Title: Dosimetric Characterization of a Dedicated Breast Computed Tomography Clinical Prototype System Purpose: To determine the radiation dosimetry characteristics of a dedicated breast computed tomography (BCT) clinical prototype system. Method and Materials: The relevant parameters of the BCT system were replicated in a Monte Carlo simulation program based on the Geant4 toolkit. To...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید


عنوان ژورنال:
journal of medical signals and sensors

جلد ۵، شماره ۲، صفحات ۱۲۳-۰

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023