Optothermal profile of an ablation catheter with integrated microcoil for MR-thermometry during Nd:YAG laser interstitial thermal therapies of the liver—an in-vitro experimental and theoretical study.

نویسندگان

  • Evdokia M Kardoulaki
  • Richard R A Syms
  • Ian R Young
  • Kaushal Choonee
  • Marc Rea
  • Wladyslaw M W Gedroyc
چکیده

PURPOSE Flexible microcoils integrated with ablation catheters can improve the temperature accuracy during local MR-thermometry in Nd:YAG laser interstitial thermal therapies. Here, the authors are concerned with obtaining a preliminary confirmation of the clinical utility of the modified catheter. They investigate whether the thin-film substrate and copper tracks of the printed coil inductor affect the symmetry of the thermal profile, and hence of the lesion produced. METHODS Transmission spectroscopy in the near infrared was performed to test for the attenuation at 1064 nm through the 25 μm thick Kapton substrate of the microcoil. The radial transmission profile of an infrared high-power, light emitting diode with >80% normalized power at 1064 nm was measured through a cross section of the modified applicator to assess the impact of the copper inductor on the optical profile. The measurements were performed in air, as well as with the applicator surrounded by two types of scattering media; crystals of NaCl and a layer of liver-mimicking gel phantom. A numerical model based on Huygens-Fresnel principle and finite element simulations, using a commercially available package (COMSOL Multiphysics), were employed to compare with the optical measurements. The impact of the modified optical profile on the thermal symmetry was assessed by examining the high resolution microcoil derived thermal maps from a Nd:YAG laser ablation performed on a liver-mimicking gel phantom. RESULTS Less than 30% attenuation through the Kapton film was verified. Shadowing behind the copper tracks was observed in air and the measured radial irradiation correlated well with the diffraction pattern calculated numerically using the Huygens-Fresnel principle. Both optical experiments and simulations, demonstrate that shadowing is mitigated by the scattering properties of a turbid medium. The microcoil derived thermal maps at the end of a Nd:YAG laser ablation performed on a gel phantom in a 3 T scanner confirm that the modified irradiation pattern does not disrupt the thermal symmetry, even though, unlike tissue, the gel is minimally scattering. CONCLUSIONS The results from this initial assessment indicate that microcoils can be safely integrated with ablation catheters and ensure that the complete necrosis of the liver tumor can still be achieved.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of Temperature Rise and Thermal Lesion Dimensions in Liver Laser Interstitial Thermotherapy

Introduction:  Laser  interstitial  thermotherapy  (LITT)  is  an  internal  ablation  therapy  method  consisting of a percutaneous or intraoperative insertion of laser fibers directly into the liver tumor with  maximum diameter of 5 cm. In this treatment method, there isn’t any general information about the  relationship  between  increasing  the  exposure  power,  coagulation  and  carboniza...

متن کامل

Gd-EOB-DTPA-Enhanced MR Guidance in Thermal Ablation of Liver Malignancies

OBJECTIVE To evaluate the potency of Gd-EOB-DTPA to support hepatic catheter placement in laser ablation procedures by quantifying time-dependent delineation effects for instrumentation and target tumor within liver parenchyma. Monitoring potential influence on online MR thermometry during the ablation procedure is a secondary aim. MATERIALS AND METHODS 30 cases of MR-guided laser ablation we...

متن کامل

Experimental Study of a Flash-lamp Pumped Passively Q-Switched Nd:YAG Laser Using Cr4+:YAG Saturable Absorber

This paper deals with the experimental results of a xenon flash-lamp pumped passively Q-switched Nd:YAG laser using Cr4+:YAG saturable absorber. The need of a laser cavity to be integrated into a time-of-flight laser range finder is of great interest as an experimental investigation of several laser resonators. Two types of laser resonator, with different lengths, have been studied: the flat-fl...

متن کامل

Feasibility of real-time MR thermal dose mapping for predicting radiofrequency ablation outcome in the myocardium in vivo

BACKGROUND Clinical treatment of cardiac arrhythmia by radiofrequency ablation (RFA) currently lacks quantitative and precise visualization of lesion formation in the myocardium during the procedure. This study aims at evaluating thermal dose (TD) imaging obtained from real-time magnetic resonance (MR) thermometry on the heart as a relevant indicator of the thermal lesion extent. METHODS MR t...

متن کامل

Numerical Study for Optimizing Parameters of High-Intensity Focused Ultrasound-Induced Thermal Field during Liver Tumor Ablation: HIFU Simulator

Introduction High intensity focused ultrasound (HIFU) is considered a noninvasive and effective technique for tumor ablation. Frequency and acoustic power are the most effective parameters for temperature distribution and the extent of tissue damage. The aim of this study was to optimize the operating transducer parameters such as frequency and input power in order to acquire suitable temperatu...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Medical physics

دوره 42 3  شماره 

صفحات  -

تاریخ انتشار 2015