Tachycardia induction due to inappropriate implantable cardioverter defibrillator therapy: what is the mechanism?

نویسندگان

  • Stylianos Tzeis
  • George Andrikopoulos
  • Ioannis Rassias
  • George Theodorakis
چکیده

Antitachycardia therapies administered by implantable cardioverter defibrillators (ICDs) can occasionally prove proarrhythmic due to induction of ventricular tachyarrhythmias. In this report we present the case of a tachycardia induction as a result of inappropriate therapy delivery in an ICD recipient. Detailed analysis of the stored electrograms facilitated the delineation of the underlying mechanism of the recorded tachycardia.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Implantable cardioverter-defibrillator in a patient with dextrocardia situs inversus

Background: Dextrocardia is a congenital anomaly, which may have coexistent coronary artery disease (CAD), arrhythmias and conventional indications for device therapy. However, the implantation of transvenous leads can be technically challenging and the approach needs to be tailored to the patient's individual anatomy. Case presentation: A 54-year-old male with dextrocardia situs inversus and i...

متن کامل

Ventricular Tachycardia Recognized as Ventricular Fibrillation by an ICD Due to Double Counting of the QRS Complex

A case of ventricular tachycardia misdiagnosed as ventricular fibrillation due to double counting of the tachycardia’s wide QRS complex in an implantable cardioverter defibrillator (ICD) recipient is presented. As a result, an inappropriate shock was delivered by the device instead of anti-tachycardia pacing. Reprogramming of the detection criteria in a less sensitive manner resolved the proble...

متن کامل

Trouble Shooting of ICD Therapy: Inappropriate Shock

Patients with an implantable cardioverter defibrillator (ICD) can undergo inappropriate therapies (shock or antitachycardia pacing) if electrical activity not originating in the ventricle is wrongly recognized as ventricular by the device. Inappropriate therapy leading to shocks is not innocuous; consequences include pain and its psychological sequelae, particularly anxiety, as well as inductio...

متن کامل

Implantable Cardioverter Defibrillator and Inappropriate Therapy: “Black Box” Examination Yielded Both Human and Technical Causes

We report on a 70-year-old male patient who was recipient of GEM III DR 7275 Cardioverter Defibrillator, and who presented with inappropriate shocks. The patient had a documented slow ventricular tachycardia (VT), and the device was programmed to detect VT at rates >100 bpm, fast VT (FVT, via VT) at rates >150 bpm, and ventricular fibrillation (VF) at rates >188 bpm. After detection of FVT, eff...

متن کامل

Outcome after implantation of a cardioverter-defibrillator in patients with Brugada syndrome: a multicenter study-part 2.

BACKGROUND Implantable cardioverter-defibrillator indications in Brugada syndrome remain controversial, especially in asymptomatic patients. Previous outcome data are limited by relatively small numbers of patients or short follow-up durations. We report the outcome of patients with Brugada syndrome implanted with an implantable cardioverter-defibrillator in a large multicenter registry. METH...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Hellenic journal of cardiology : HJC = Hellenike kardiologike epitheorese

دوره 51 6  شماره 

صفحات  -

تاریخ انتشار 2010