Reproducibility of aortic intima-media thickness in infants using edge-detection software and manual caliper measurements

نویسندگان

  • Kate McCloskey
  • Anne-Louise Ponsonby
  • John B Carlin
  • Kim Jachno
  • Michael Cheung
  • Michael R Skilton
  • Jane Koleff
  • Peter Vuillermin
  • David Burgner
چکیده

BACKGROUND Aortic intima-media thickness measured by transabdominal ultrasound (aIMT) is an intermediate phenotype of cardiovascular risk. We aimed to (1) investigate the reproducibility of aIMT in a population-derived cohort of infants; (2) establish the distribution of aIMT in early infancy; (3) compare measurement by edge-detection software to that by manual sonographic calipers; and (4) assess the effect of individual and environmental variables on image quality. METHODS Participants were term infants recruited to a population-derived birth cohort study. Transabdominal ultrasound was performed at six weeks of age by one of two trained operators. Thirty participants had ultrasounds performed by both operators on the same day. Data were collected on environmental (infant sleeping, presence of a sibling, use of sucrose, timing during study visit) and individual (post-conception age, weight, gender) variables. Two readers assessed image quality and measured aIMT by edge-detection software and a subset by manual sonographic calipers. Measurements were repeated by the same reader and between readers to obtain intra-observer and inter-observer reliability. RESULTS Aortic IMT was measured successfully using edge-detection in 814 infants, and 290 of these infants also had aIMT measured using manual sonographic calipers. The intra-reader intra-class correlation (ICC) (n = 20) was 0.90 (95% CI 0.76, 0.96), mean difference 1.5 μm (95% LOA -39, 59). The between reader ICC using edge-detection (n = 20) was 0.92 (95% CI 0.82, 0.97) mean difference 2 μm (95% LOA -45.0, 49.0) and with manual caliper measurement (n = 290) the ICC was 0.84 (95% CI 0.80, 0.87) mean difference 5 μm (95% LOA -51.8, 61.8). Edge-detection measurements were greater than those from manual sonographic calipers (mean aIMT 618 μm (50) versus mean aIMT 563 μm (49) respectively; p < 0.001, mean difference 44 μm, 95% LOA -54, 142). With the exception of infant crying (p = 0.001), no associations were observed between individual and environmental variables and image quality. CONCLUSION In a population-derived cohort of term infants, aIMT measurement has a high level of intra and inter-reader reproducibility. Measurement of aIMT using edge-detection software gives higher inter-reader ICC than manual sonographic calipers. Image quality is not substantially affected by individual and environmental factors.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Ghrelin Does not Alter Aortic Intima-Media Thickness and Adipose Tissue Characteristics in Control and Obese Mice

Objective(s): Atherosclerosis is a chronic immune-inflammatory disease that generally leads to ischemic heart disease. Ghrelin has several modulatory effects on cardiovascular system. In this study, we investigated the effect of ghrelin on aortic intima-media thickness, size and the number of adipocyte cells in obese and control mice. Materials and Methods:This study was conducted on 24 male C...

متن کامل

Effect of blockade of neuropeptide Y receptor on aortic intima-media thickness and adipose tissue characteristics in normal and obese mice

Objective(s): Atherosclerosis is an important risk factor for coronary heart disease. Neuropeptide Y (NPY) and its receptors, located in peripheral tissue such as white adipose tissue, have been linked to obesity and fat storage. The role of NPY in atherosclerosis has not yet been fully studied, so this study was conducted to further investigate the effect of BIIE 0246, an NPY receptor antagoni...

متن کامل

Associations of edge-detected and manual-traced common carotid intima-media thickness measurements with Framingham risk factors: the multi-ethnic study of atherosclerosis.

BACKGROUND AND PURPOSE Carotid intima-media thickness (IMT) is a marker of cardiovascular disease derived from ultrasound images of the carotid artery. In most outcome studies, human readers identify and trace the key IMT interfaces. We evaluate an alternate approach using automated edge detection. METHODS We studied a subset of 5640 participants with an average age 61.7 years (48% men) of th...

متن کامل

Comparison of Abdominal Aorta Intima-Media Thickness, Serum Level of Leptin, and Lipid Profile between Maternal Preeclampsia and Neonatal Outcomes

Background Pre-eclampsia is one of the main causes of morbidity and mortality among pregnant women. This study was done with the objective of measuring abdominal aorta intima-media thickness, serum level of leptin, and lipid profile in neonates of pre-eclampsia mothers and comparing them with neonates of mothers without pre-eclampsia (control group). Materials and Methods Ninety neonates of pre...

متن کامل

Carotid artery Intima-Media thickness and cardiovascular risk factors in chronic hemodialysis patients

Introduction: Common carotid artery intima-media thickness (CCA-IMT) correlates with cardiovascular mortality in patients under hemodialysis. In this investigation, we evaluated relationship between CCA-IMT with dependent and independent cardiovascular risk factors in chronic hemodialysis patients. Methods: In this cross-sectional study, 51 chronic hemodialysis patients, (31 males and 20 fema...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 12  شماره 

صفحات  -

تاریخ انتشار 2014