Paper I Experimental validation of a method for removing the capacitive leakage artifact from electrical bioimpedance spectroscopy measurements
نویسندگان
چکیده
Often when performing electrical bioimpedance (EBI) spectroscopy measurements, the obtained EBI data present a hook-like deviation, which is most noticeable at high frequencies in the impedance plane. The deviation is due to a capacitive leakage effect caused by the presence of stray capacitances. In addition to the data deviation being remarkably noticeable at high frequencies in the phase and the reactance spectra, the measured EBI is also altered in the resistance and the modulus. If this EBI data deviation is not properly removed, it interferes with subsequent data analysis processes, especially with Cole model-based analyses. In other words, to perform any accurate analysis of the EBI spectroscopy data, the hook deviation must be properly removed. Td compensation is a method used to compensate the hook deviation present in EBI data; it consists of multiplying the obtained spectrum, Zmeas(ω), by a complex exponential in the form of exp(–jωTd). Although the method is well known and accepted, Td compensation cannot entirely correct the hook-like deviation; moreover, it lacks solid scientific grounds. In this work, the Td compensation method is revisited, and it is shown that it should not be used to correct the effect of a capacitive leakage; furthermore, a more developed approach for correcting the hook deviation caused by the capacitive leakage is proposed. The method includes a novel correcting expression and a process for selecting the proper values of expressions that are complex and frequency dependent. The correctness of the novel method is validated with the experimental data obtained from measurements from three different EBI applications. The obtained results confirm the sufficiency and feasibility of the correcting method.
منابع مشابه
Experimental validation of a method for removing the capacitive leakage artifact
Often when performing electrical bioimpedance (EBI) spectroscopy measurements, the obtained EBI data present a hook-like deviation, which is most noticeable at high frequencies in the impedance plane. The deviation is due to a capacitive leakage effect caused by the presence of stray capacitances. In addition to the data deviation being remarkably noticeable at high frequencies in the phase and...
متن کاملA Novel Approach for Removing the Hook Effect Artefact from Electrical Bioimpedance Spectroscopy Measurements
Very often in Electrical Bioimpedance (EBI) spectroscopy measurements the presence of stray capacitances creates a measurement artefact commonly known as Hook Effect. Such an artefact creates a hook-alike deviation of the EBI data noticeable when representing the measurement on the impedance plane. Such Hook Effect is noticeable at high frequencies but it also causes a data deviation at lower m...
متن کاملDetection and Classification of Measurement Errors in Bioimpedance Spectroscopy
Bioimpedance spectroscopy (BIS) measurement errors may be caused by parasitic stray capacitance, impedance mismatch, cross-talking or their very likely combination. An accurate detection and identification is of extreme importance for further analysis because in some cases and for some applications, certain measurement artifacts can be corrected, minimized or even avoided. In this paper we pres...
متن کاملElectrical Bioimpedance Based Monitoring of Cardiac and Respiratory Activity: Positioning of Capacitive Electrodes for Wearable Human Monitoring Device
In this paper, some observations concerning the capacitive measurements of respiratory and cardiac activity of human by using a wearable electrode shirt, are discussed. Electrode shirt is described and the results of the measurements are presented together with the charts of the signals. The related aspects, originating from the observations, concerning the measurement results and the design of...
متن کاملRemoving ECG Artifact from the Surface EMG Signal Using Adaptive Subtraction Technique
Background: The electrocardiogram artifact is a major contamination in the electromyogram signals when electromyogram signal is recorded from upper trunk muscles and because of that the contaminated electromyogram is not useful.Objective: Removing electrocardiogram contamination from electromyogram signals.Methods: In this paper, the clean electromyogram signal, electrocardiogram artifact and e...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -
تاریخ انتشار 2012