Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates

نویسندگان

  • Christopher Ioannou
  • Matthew Knight
  • Luca Daniele
  • Lee Flueckiger
  • Ezekiel S L Tan
چکیده

BACKGROUND The objective of this study is to analyse the effectiveness of the surgical torque limiter during operative use. The study also investigates the potential differences in torque between hand and drill-based screw insertion into locking plates using a standardised torque limiter. METHODS Torque for both hand and power screw insertion was measured through a load cell, registering 6.66 points per second. This was performed in a controlled environment using synthetic bone, a locking plate and locking screws to simulate plate fixation. Screws were inserted by hand and by drill with torque values measured. RESULTS The surgical torque limiter (1.5 Nm) was effective as the highest recorded reading in the study was 1.409 Nm. Comparatively, there is a statistically significant difference between screw insertion methods. Torque produced for manually driven screw insertion into locking plates was 1.289 Nm (95 % CI 1.269-1.308) with drill-powered screw insertion at 0.740 Nm (95 % CI 0.723-0.757). CONCLUSIONS The surgical torque limiter proved to be effective as per product specifications. Screws inserted under power produce significantly less torque when compared to manual insertion by hand. This is likely related to the mechanism of the torque limiter when being used at higher speeds for which it was designed. We conclude that screws may be inserted using power to the plate with the addition of a torque limiter. It is recommended that all screws inserted by drill be hand tightened to achieve adequate torque values.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A Pitfall in Fixation of Distal Humeral Fractures with Pre-Contoured Locking Compression Plate

Anatomically precontoured locking plates are intended to facilitate the fixation of articular fractures and particularly those associated with osteoporosis. Fractures of the distal humerus are relatively uncommon injuries where operative intervention can be exceptionally challenging. The distal humeral trochlea provides a very narrow anatomical window through which to pass a fixed-angle locking...

متن کامل

The effect of screw angulation and insertion torque on the push-out strength of polyaxial locking screws and the single cycle to failure in bending of polyaxial locking plates.

OBJECTIVE To evaluate the mechanical properties of the Polyaxial Advanced Locking System (PAX) in screw push-out and four-point bending. MATERIALS AND METHODS Screw push-out: PAX locking screws were applied to first generation PAX plates at three different insertion angles with two different insertion torques. A load was applied parallel to the screw axis, and screw push-out force was measure...

متن کامل

Torque Removal Evaluation of Screw in One-Piece and Two-Piece Abutments Tightened with a Handheld screwdriver

Introduction: Some clinicians use a handheld screw driver instead of a torque wrench to definitively tighten abutment screws. The aim of this study was to compare the removal torque of one-piece and two-piece abutments tightened with a handheld driver and a torque control ratchet. Methods: 40 ITI implants were placed in acrylic blocks and divided into 4 groups. In groups one and two, 10 ITI one...

متن کامل

Evaluation of insertion torque of different orthodontic miniscrews: An in vitro study

*Department of P.O.P, college of dentistry, university of Mosul, Mosul, Iraq. Introduction Conservation of anchorage in totality has been a perennial problem to the traditional orthodontist. Conventional means of supporting anchorage have been used by either intra oral sites or relying on extra oral means. Both of these have their limitations. The extra oral forces very difficult device and som...

متن کامل

Automated surgical screwdriver: automated screw placement.

The use of power screwdrivers and drills for tapping and screw insertion in surgery is becoming more common. It has been established from clinical observations that the use of a small air drill for inserting self-tapping screws provides improved coaxial alignment and precision, and that the drill should be stopped before the screw head is completely seated on the plate, presumably to reduce the...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 11  شماره 

صفحات  -

تاریخ انتشار 2016