Efficacy and temporal stability of reduced safety margins for ventricular defibrillation: primary results from the Low Energy Safety Study (LESS).
نویسندگان
چکیده
BACKGROUND Traditionally, a safety margin of at least 10 J between the maximum output of the pulse generator and the energy needed for ventricular defibrillation has been used because lower safety margins were associated with unacceptably high rates of failed defibrillation and sudden cardiac death. The Low Energy Safety Study (LESS) was a prospective, randomized assessment of the safety margin requirements for modern implantable cardioverter-defibrillator (ICD) systems. METHODS AND RESULTS A total of 636 patients undergoing initial ICD implantation with a dual-coil lead and active pulse generator were evaluated. The defibrillation threshold (DFT) and enhanced DFT (DFT+ and DFT++) were measured using a modified step-down protocol. Conversion testing of induced ventricular fibrillation before discharge, at 3 months, and at 12 months was performed, as was randomization to chronic programming at either 2 steps above DFT++ or maximal output. The induced ventricular fibrillation data had conversion success rates of 91.4%, 97.9%, 99.1%, 99.6%, and 99.8% for safety margins of 0, 1, 2, 3, and 4 steps above the DFT++, respectively. A margin of 4 to 6 J was adequate to maintain high conversion success over time (98.9% before discharge versus 99.2% at 12 months; P=NS). Over a mean follow-up of 24+/-13 months, conversion of spontaneously occurring ventricular tachyarrhythmias >200 bpm was identical (97.3%), despite a safety margin difference of 5.2+/-1.1 J for the 2-step group versus 20.8+/-4.2 J for maximal output. CONCLUSIONS With a rigorous implantation algorithm, a safety margin of about 5 J is adequate for safe implantation of modern ICD systems.
منابع مشابه
Efficacy and Temporal Stability of Reduced Safety Margins for Ventricular Defibrillation
Background—Traditionally, a safety margin of at least 10 J between the maximum output of the pulse generator and the energy needed for ventricular defibrillation has been used because lower safety margins were associated with unacceptably high rates of failed defibrillation and sudden cardiac death. The Low Energy Safety Study (LESS) was a prospective, randomized assessment of the safety margin...
متن کاملA prospective evaluation of two defibrillation safety margin techniques in patients with low defibrillation energy requirements.
INTRODUCTION In patients undergoing defibrillator implantation, an appropriate defibrillation safety margin has been considered to be either 10 J or an energy equal to the defibrillation energy requirement. However, a previous clinical report suggested that a larger safety margin may be required in patients with a low defibrillation energy requirement. Therefore, the purpose of this prospective...
متن کاملRelative Stability Improves
This study tested the hypothesis that subjects improve their relative stability as they learn a dynamic pulling task. Healthy adult subjects practiced making brief horizontal pulls (<300 ms) on a handle to a range of target forces ranging from 20 to 80 percent of their estimated maximum for 5 days. They were instructed to always keep their feet flat, and begin and end their motion in an upright...
متن کاملAutomated vulnerability testing identifies patients with inadequate defibrillation safety margin.
BACKGROUND Implantable cardioverter-defibrillator system efficacy is tested at implant by induction of ventricular fibrillation (VF). Defibrillation safety margin can be assessed without VF induction using upper limit of vulnerability methods, but these methods have required manual determination of T-wave timing. METHODS AND RESULTS To test the feasibility of an inductionless system of implan...
متن کاملTemporal stability and precision of ventricular defibrillation threshold data.
Over 200 measurements of the minimum damped sinusoidal current and energy for transchest electrical ventricular defibrillation (ventricular defibrillation threshold) were made to determine the stability and precision of threshold data in 15 pentobarbital-anesthetized dogs. Threshold was determined by repeated trials of fibrillation and defibrillation with successive shocks of diminishing curren...