The Anaconda AAA stent graft system: 2-year clinical and technical results of a multicentre clinical evaluation.

نویسندگان

  • S G J Rödel
  • R H Geelkerken
  • R J Prescott
  • H J Florek
  • P Kasprzak
  • J Brunkwall
چکیده

INTRODUCTION This study reports the technical and mid-term clinical results of the second-generation Anaconda AAA Stent Graft System endovascular device for treatment of abdominal aortic aneurysm (AAA). The design of the Anaconda AAA Stent Graft System is characterised by a three-piece system consisting of two proximal independent saddle-shaped nitinol self-expandable rings with hooks fixation, zero body support and vacuum-cleaner tube leg design. METHODS From July 2002 to April 2005, a total of 61 patients with AAA were enrolled in a multicentre, prospective, non-randomised controlled design study. All patients received a second-generation Anaconda AAA Stent Graft System. They entered a standard follow-up protocol at discharge for 3, 6, 12 and 24 months. Follow-up data included survival; rupture-free survival; incidence of aneurysm rupture, death from aneurysm rupture, aneurysm-related death; freedom from aneurysm expansion; freedom from Types I and III endoleaks; endograft patency and technical and clinical success rates. RESULTS Successful access to the arterial system was achieved in all patients. The primary technical success was 59 out of 61 and the primary assisted technical success was 60 out of 61. All endovascular grafts were patent without significant twists, kinks or obstructions. Migration was not observed in any of the grafts. During the first 30-day period, two serious adverse events (3%), both not related to the procedure, were observed. Nine patients (15%) needed a secondary intervention; two of these interventions were related to stent graft (3%). The mean aneurysm sac diameter decreased significantly from 57 mm pre-operative to 45mm after 24 months, without aneurysm growth. There was one Type I endoleak at initial implantation, which was corrected using a proximal extension cuff. In total, three Type II endoleaks were still present after 24 months without any signs of aneurysm growth. CONCLUSION The design features of the second-generation Anaconda AAA Stent Graft System are effective in the treatment of AAAs on mid-term evaluation.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Results of the Anaconda endovascular graft in abdominal aortic aneurysm with a severe angulated infrarenal neck.

OBJECTIVE Proximal neck anatomy of an abdominal aortic aneurysm (AAA), especially a severe angulated neck of more than 60 degrees, predicts adverse outcome in endovascular aneurysm repair. In the present study, we evaluate the feasibility of the use of the Anaconda endovascular graft (Vascutec, Terumo, Inchinnan, Scotland) for treating infrarenal AAA with a severe angulated neck (>60 degrees) a...

متن کامل

Management of a large coronary artery aneurysm with graft-coated stent during the acute phase of myocardial infarction

Coronary aneurysms in adults are rare clinical entities. Herein, we presented a 54-year-old man who was admitted with an acute extensive anterior myocardial infarction. The emergency coronary angiography revealed an isolated large aneurysm at the proximal segment of the left anterior descending coronary artery. The patient was successfully treated with a graft-coated stent

متن کامل

Anaconda Experiment in Endovascular Stent Therapy of Abdominal Aortic Aneurysms: Early Results of a Single Center

Background: We investigated the efficacy, preoperative, postoperative and early stage outcomes of Anaconda type endovascular stent-grafts in patients who had intervention in our clinic with the diagnosis of abdominal aortic aneurysm. Material – Methods: During the years of 2010 and 2013, 23 patients (19 male, 4 female) had endovascular stent-graft procedure with the diagnosis of infrarenal abdo...

متن کامل

The new C3 Gore Excluder stent-graft: single-center experience with 100 patients.

OBJECTIVES To present results from the first 100 patients treated with the new C3 Gore Excluder stent-graft in a single institution. METHODS All patients treated with the C3 Excluder stent-graft between August 2010 and July 2013 in our institution were included. Patient demographics, treatment indication, need for intraoperative stent-graft repositioning, immediate technical success, survival...

متن کامل

Endovascular Treatment of Isolated Iliac Artery Aneurysms with Anaconda Stent Graft Limb

Isolated aneurysms of the iliac arteries are relatively rare conditions that traditionally have been treated by surgical reconstruction. We report our experience with endovascular treatment of iliac artery aneurysms (IAAs) with Anaconda stent graft limb. Two male patients were found to have 4.5 and 3.6 cm isolated common IAAs, respectively. The endograft was successfully advanced and deployed p...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery

دوره 38 6  شماره 

صفحات  -

تاریخ انتشار 2009