Kinetics of defibrillation shock-induced response: design implications for the optimal defibrillation waveform.
نویسنده
چکیده
INTRODUCTION Implantable cardioverter defibrillator (ICD) therapy is a well-established therapy for treating patients at high risk for sudden cardiac death. Recently formulated virtual electrode polarization theory is a promising foundation for the theory of defibrillation. Yet, continuing optimization of defibrillation therapy is limited to primarily empirical methods due to difficulties in assessing kinetics of cellular response in whole heart models of defibrillation. The aim of this study was to evaluate the response of the myocardium in the context of virtual electrode polarization. METHODS AND RESULTS We used a Langendorff-perfused rabbit heart model of ICD therapy and voltage-sensitive fluorescent dye imaging in order to map kinetics of trans membrane potential during both mono- and biphasic shocks applied at various phases of the QT-interval. Cellular response was fitted to a single exponential function using the Levenberg-Marquardt method. Time constants (tau) were measured in 45 288 optical records from 17 hearts. We found that cellular response depends upon both QT-phase of application, intensity, polarity, and phase of the biphasic waveform. Shocks of larger strengths produce a faster response. The tau of the first-phase negatively polarizing response was significantly larger compared with the positively polarizing response at intensities below 200 V, but smaller at 200 V and above. The tau of the second phase negatively polarizing response was always slower than the positively polarizing response, regardless of amplitude, and timing. Overall, tau ranged from 1.6 ms to 14.2 ms. CONCLUSIONS The time constant of the membrane depends on the field, action potential phase and the shock polarity, but exceeds 1 msec. Therefore, we suggest using a slower shock leading edge, since the membrane cannot follow potentially damaging faster waveforms.
منابع مشابه
Mechanisms of Pacing and Defibrillation
Kinetics of defibrillation shock-induced response: design implications for the optimal defibrillation waveform Mowrey and colleagues investigated the response of the myocardium with respect to defibrillation shocks . However, their trials are concerned with effective (Fig. 10) or partially ineffective (Fig. 6) voltage pulses within or outside the absolute refractory period. This is remarkable a...
متن کاملOptimization of biphasic waveforms for human nonthoracotomy defibrillation.
BACKGROUND Biphasic waveforms reduce defibrillation threshold (DFT) in a wide variety of models. Although there are several human studies of long-duration, high-tilt biphasic waveform defibrillation, the specific biphasic waveform shape required to achieve optimal DFT reduction is unknown. METHODS AND RESULTS This study tested the effect of single capacitor biphasic waveform tilt modification...
متن کاملImproved safety factor for triphasic defibrillator waveforms.
Newly developed biphasic waveforms improve defibrillation efficacy both by reduction of defibrillation threshold and by amelioration of shock-induced dysfunction depending on the relative shape of the first and second pulses. Each of these independent effects improves the waveform's safety factor, the ratio between the shock intensity that produces a specific degree of postshock dysfunction and...
متن کاملIs there an Optimal Shape of the Defibrillation Shock: Constant Current vs. Pulsed Biphasic Waveforms?
Three waveforms for transthoracic defibrillation are assessed and compared: the Pulsed Biphasic Waveform (PBW), the Rectilinear Biphasic Waveform (RBW), and the “lossless” constant current (LLCC) pulses. Two indices are introduced: 1) kf = W/W0 – the ratio between the delivered energy W and the energy W0 of a rectangular pulse with the same duration and electric charge; 2) ηC = W/WC0 – the leve...
متن کاملComparison of a novel rectilinear biphasic waveform with a damped sine wave monophasic waveform for transthoracic ventricular defibrillation. ZOLL Investigators.
OBJECTIVES We compared the efficacy of a novel rectilinear biphasic waveform, consisting of a constant current first phase, with a damped sine wave monophasic waveform during transthoracic defibrillation. BACKGROUND Multiple studies have shown that for endocardial defibrillation, biphasic waveforms have a greater efficacy than monophasic waveforms. More recently, a 130-J truncated exponential...