Which apparatus for inhaled pentamidine? A comparison of pulmonary deposition via eight nebulisers.
نویسنده
چکیده
Aerosolised pentamidine 300 mg in 5 or 6 ml solution was administered via 8 different nebuliser systems to 12 patients with acquired immunodeficiency syndrome. Using 99mTc human serum albumin as an indirect marker for pentamidine, pulmonary, extrapulmonary (gastric and oropharyngeal) and alveolar deposition of pentamidine were measured using a gamma camera. Side effects (visual analogue scales) and changes in lung function associated with each treatment were also quantified. Deposition was completed more rapidly with the ultrasonic than the jet nebulisers. Mean total pulmonary depositions (mg +/- SEM) were Respirgard II, 6.1 +/- 0.5; Centimist, 7.3 +/- 1.0, System 22 Mizer, 14.3 +/- 2.1; System 22 Mizer with particle separator; 4.5 +/- 0.4; System 22 Mizer with Optimist 2, 6.3 +/- 0.9; Fisoneb, 6.0 +/- 1.2; Pentasonic (Portasonic); 4.6 +/- 0.9; Samsonic, 2.9 +/- 0.4. Differences between the nebulisers for peripheral lung and alveolar deposition reflected this pattern. Side effects scores were largest with System 22 Mizer, Pentasonic (Portasonic), and Fisoneb, and these produced the greatest oropharyngeal and gastric deposition. The largest reductions in lung function were associated with System 22 Mizer. A 300 mg dose of pentamidine nebulised via Respirgard II is known to be effective prophylaxis for Pneumocystis carinii pneumonia when given once monthly. Our results show that equivalent pulmonary deposition can be produced by other nebulisers. System 22 Mizer gives over twice the deposition associated with Respirgard II, and used with a pentamidine dose of 150 mg is likely to produce an adequate lung dose for prophylaxis. This nebuliser, however, is associated with more marked side effects.
منابع مشابه
Pulmonary deposition of nebulised pentamidine isethionate: effect of nebuliser type, dose, and volume of fill.
An estimate of the absolute pulmonary deposition of nebulised pentamidine isethionate was obtained in nine patients with AIDS. Two nebuliser systems were compared, System 22 Mizer (Medic-Aid) and Respirgard II (Marquest), with 50 and 150 mg doses of pentamidine in a 3 ml solution driven by an air flow of 6 l/min with the patient in the sitting position. The 50 mg pentamidine dose was repeated w...
متن کاملNebulisers for patients with HIV infection and AIDS.
bacterium tuberculosis and other opportunistic infections. The choice of nebuliser and delivery system is If nebulisers are to be used within the hospital dictated by the need to deposit pentamidine then nebulisation should be carried out in a in the alveoli and to avoid deposition in the room separate from the main ward area, ideally oropharynx, trachea, and major bronchi. ...
متن کاملPulmonary deposition of nebulised amiloride in cystic fibrosis: comparison of two nebulisers.
BACKGROUND Preliminary evidence suggests that regular inhalation of nebulised amiloride reduces sputum viscoelasticity, increases the clearance of sputum by mucociliary mechanisms and by coughing and reduces the rate of deterioration in lung function in patients with cystic fibrosis. These effects depend on adequate delivery of amiloride to the airways. This study was performed to quantify and ...
متن کاملInhaled Pentamidine Prophylaxis in Lung Transplant R#{233}cipients*
The incidence ofPneumocystis carinu pneun onia (PCP) has b en shown to be high posttransplantation in the absence ofprophylaxis. For this reason, lung transplant recipients routinely receive prophylaxis. We report on our results using aerosolized pentamidine prophylaxis in nine patients post-lung transplantation (eight single lung transplants, one double). The patients received monthly treatmen...
متن کاملNebulised pentamidine as treatment for Pneumocystis carinii pneumonia in the acquired immunodeficiency syndrome.
Nebulised pentamidine was used to treat 30 patients with Pneumocystis carinii pneumonia. Fourteen patients (group 1) received pentamidine isethionate 4 mg/kg (six patients) or 8 mg/kg (eight patients) via a standard jet nebuliser (Acorn, system 22) with a flow rate of 8 l/min. The aerosol droplets had a mass median aerodynamic diameter of 2.6 microns (geometric standard deviation (GSD) 2.9) and...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- The European respiratory journal
دوره 4 5 شماره
صفحات -
تاریخ انتشار 1991