Evaluation of 16 adult disposable manual resuscitators.

نویسندگان

  • Daniel G Mazzolini
  • Nancy A Marshall
چکیده

INTRODUCTION Disposable manual resuscitators are commonly used to ventilate patients during cardiopulmonary resuscitation, suctioning, and intrahospital transport, and their clinical performance is critical. METHODS We bench-tested 16 adult disposable manual resuscitators from 9 different manufacturers. We performed a series of tests and made observations using testing industry standards as a guideline. Each resuscitator was tested for fraction of delivered oxygen (FDO2), tidal volume delivery, drop test, and patient valve lock-up. We also made observations about reservoir style, ease or difficulty of attaching the positive end-expiratory pressure valve, size, texture, carbon dioxide detector, and if the resuscitator was labeled "latex free." RESULTS Reservoir style and manufacturer design significantly affected FDO2. In general, the resuscitators with reservoir bags provided better FDO2 than did the resuscitators with tubing reservoirs (large-bore or small-bore). Delivered tidal volumes were acceptable for all the resuscitators tested. All the resuscitators passed a standard drop test. None of the resuscitators had a patient valve lock up at high flow. With all but one resuscitator, attaching the positive end-expiratory pressure valve was easy and the valve attached securely. Most resuscitators were average in size and had good texture, but some were large, somewhat slippery, and difficult to handle. Only 2 resuscitators came with carbon dioxide detectors already attached. All but one of the resuscitators were labeled "latex free," and the one that was not was found not to contain latex proteins. CONCLUSIONS Resuscitator reservoir style and manufacturer design significantly affect FDO2. Some resuscitator models may not deliver adequate oxygen in certain clinical circumstances. Each institution should evaluate and choose the resuscitator that best fits its needs, while meeting established performance criteria.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Efficiency of manual resuscitators at birth.

The effectiveness of five neonatal/paediatric manual resuscitators was assessed in a group of babies born by caesarean section. Results showed that devices incorporating a large volume reservoir produced the greatest tidal volume, while those with smaller volume reservoirs could not be considered satisfactory for routine use during neonatal resuscitation.

متن کامل

Evaluation of manual resuscitators used in ICUs in Brazil*

OBJECTIVE To evaluate the performance of manual resuscitators (MRs) used in Brazil in accordance with international standards. METHODS Using a respiratory system simulator, four volunteer physiotherapists employed eight MRs (five produced in Brazil and three produced abroad), which were tested for inspiratory and expiratory resistance of the patient valve; functioning of the pressure-limiting...

متن کامل

Performance and Acceptability of Two Self-Inflating Bag-Mask Neonatal Resuscitator Designs.

BACKGROUND A self-inflating bag-mask device is specified by international policy guidelines as standard prototype of care for newborn resuscitation. Our hypothesis is that a new bag-mask design would be as effective and easy to use as a standard, self-inflating resuscitation bag-mask. METHODS We conducted a comparative evaluation of the performance and acceptability of the Laerdal 220-mL resu...

متن کامل

Evaluation of Ecological Factors Regarding the Use of Disposable Containers

 Introduction: Excessive consumption of disposable containers, and specifically the plastic ones, has adverse effects on ecosystem and the community's health. To identify ecological factors (individual, interpersonal and social) related to these containers, this study aimed to specify the most important factors according to the participants. Methods: This was a qualitative content analysis  co...

متن کامل

[Failure of self-inflating manual resuscitator due to the presence of dried pulmonary secretions].

BACKGROUND AND OBJECTIVES Self-inflating manual resuscitators (SIMR) are devices used to ventilate patients with ventilatory needs. These devices consist of a set of valves that work sequentially, and changes in their function may be harmful to patients. CASE REPORT During the use of SIMR, it was observed that the valve remained fixed due to the presence of dried pulmonary secretion, which ma...

متن کامل