Measurement of urinary chloride with the Kodak Ektachem 400.

نویسندگان

  • M F Melhem
  • H Seltman
  • A Sanghvi
چکیده

Urinary chloride can be measured simply, precisely, and accurately with the Ektachem 400 clinical chemistry analyzer. Because the coated film technique used in this instrument requires the presence of a protein matrix, we added a protein-based calibrator to the urine samples to overcome the low protein concentration in these specimens. Its advantages over a coulometric titration method (Astra analyzer) include the extension of the range of linearity to lower concentrations and the absence of interference from prednisone metabolites in urine. A disadvantage is the negative bias interference of allopurinol.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Factitiously high sodium activities on the Ektachem 400 owing to interferences by high gamma-globulin concentrations.

Sera from patients with above-normal gamma-globulin concentrations give sodium values as much as 10% higher when measured in the Kodak Ektachem 400 analyzer than results obtained with the Beckman Astra-8 analyzer. Results obtained with the Astra were in agreement with flame photometry. A similar relative bias was observed for potassium measurement in the Ektachem. This discrepancy was corrected...

متن کامل

Clinical evaluation of Eastman Kodak's Ektachem 400 Analyzer.

We evaluated the performance of the Kodak Ektachem 400 Analyzer in a 16-week clinical trial. We assessed four potentiometric tests and nine colorimetric tests for precision and correlation with results obtained some other commonly used instruments (Technicon SMA II and C800 System, Du Pont aca II, and Baker CentrifiChem). The comparison was favorable for all tests except albumin, sodium, and ca...

متن کامل

Discrepancies between sodium concentrations measured by the Kodak Ektachem 700 and by dilutional and direct ion-selective electrode analyzers.

We have identified rare (approximately 0.2% of all samples), but clinically significant, discrepancies between serum or plasma sodium concentrations measured with the Kodak Ektachem 700's direct ion-selective electrode (ISE) method and concentrations measured with two other analyzers: the Beckman Synchron CX3's dilutional ISE instrument and the Radiometer KNA2 instrument for sodium-potassium an...

متن کامل

Methotrexate interferes with determinations of conjugated bilirubin with the Kodak Ektachem 400.

Immediately after intravenous infusion of a high dose (concentration in serum greater than 1000 mumol/L) of methotrexate, the apparent conjugated bilirubin (Bc) concentrations in serum of two osteosarcoma patients, as measured by the Kodak Ektachem 400 analyzer, were greater than the corresponding total bilirubin concentrations, but decreased as the concentrations of methotrexate in serum decre...

متن کامل

Benzalkonium interference with test methods for potassium and sodium.

Six automated instruments that measure sodium and potassium were tested for interference from two compounds used in catheters. Tridodecylmethylammonium heparin did not interfere with any of the methods. However, benzalkonium heparin falsely increased sodium measurement with the Kodak Ektachem, and falsely increased potassium measurements with three instruments (Beckman Astra, Baxter Paramax, an...

متن کامل