Comparison of Combat Gauze and TraumaStat in two severe groin injury models.

نویسندگان

  • Françoise Arnaud
  • Kohsuke Teranishi
  • Tomoaki Okada
  • Dione Parreño-Sacdalan
  • Daniel Hupalo
  • George McNamee
  • Walter Carr
  • David Burris
  • Richard McCarron
چکیده

BACKGROUND Fabric-like hemostatic dressings offer promise for hemorrhage control in noncompressible areas, especially given their similarity in form to standard gauze currently in use. Recently, two such products, Combat Gauze (CBG) and TraumaStat (TMS), were introduced. Their performance is evaluated in two vascular injury models. MATERIALS AND METHODS The dressings were evaluated in anesthetized Yorkshire pigs, hemorrhaged by full transection of the femoral vasculature with 2 min free bleeding period (CBG = 6, TMS = 6) or by 4 mm femoral arterial puncture with 45 s free bleeding period (CBG = 8, TMS = 8). After injury, dressings were applied, followed by 5 min of manual compression and then 500 mL resuscitation fluid infused over 30 min. Vital signs, blood pressure, and blood loss were recorded throughout the 3-h experiment. Bleeding control was the primary outcome. RESULTS All animals had similar pretreatment mean arterial pressure (MAP) (∼ 36.5 mmHg); pretreatment blood loss following injury was similar for both dressing groups in the two models [24% ± 8% estimated blood volume (EBV) 2 min after transection and 17% ± 4% EBV 45 s after puncture. Incidence of post-treatment bleeding, primarily occurring after release of manual compression or restoration of blood pressure, was more frequent in the puncture model (17% with both CBG and TMS) than the transection model (57% with CBG versus 75% with TMS). Post-treatment blood loss not controlled by the dressing was 19% ± 22% and 31% ± 17% EBV, for CBG and TMS, respectively. Survival rate was 100% for both dressings in the transection model, and was 88% for CBG and 50% for TMS in the puncture model. CONCLUSIONS These findings indicated that CBG and TMS were similarly effective in improving hemostasis. These two fabric-like dressings showed easy application and removal, leaving a clean wound for surgical repair.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A novel highly porous silica and chitosan-based hemostatic dressing is superior to HemCon and gauze sponges.

BACKGROUND Hemostatic dressings have become increasingly popular as the optimal initial treatment for severe hemorrhage. The purpose of this study was to compare the hemostatic properties of a novel highly porous silica and chitosan-based dressing (TraumaStat) to HemCon, and gauze dressing in a severe groin injury model in swine. METHODS Thirty swine were blindly randomized to receive TraumaS...

متن کامل

An evidence-based review of the use of a combat gauze (QuikClot) for hemorrhage control.

Trauma is a leading cause of morbidity and mortality. Uncontrolled hemorrhage related to the traumatic event is often the major cause of complications and death. The use of hemostatic agents may be one of the easiest and most effective methods of treating hemorrhage. The US military recommends a hemostatic combat gauze (QuikClot Combat Gauze) as the first-line hemostatic agent for use in treatm...

متن کامل

Determination of efficacy of novel modified chitosan sponge dressing in a lethal arterial injury model in swine.

BACKGROUND Chitosan is a functional biopolymer that has been widely used as a hemostat. Recently, its efficacy has been questioned due to clinical failures as a result of poor adhesiveness. The purpose of this study was to compare, in a severe groin injury model in swine, the hemostatic properties of an unmodified standard chitosan sponge with standard gauze dressing and a novel hydrophobically...

متن کامل

Comparison of Celox Gauze haemostat with kaolin based gauze

Objectives: Uncontrolled hemorrhage remains a leading cause of traumatic death. Several topicaladjunct agents have been shown to be effective in controlling hemorrhage, and two, chitosan waferdressing (HemCon [HC]) and zeolite powder dressing (QuikClot [QC]), are being utilized regularly on thebattlefield. However, recent literature reviews have concluded that no ideal topical agent...

متن کامل

Hemostatic Efficiency and Wound Healing Properties of Natural Zeolite Granules in a Lethal Rabbit Model of Complex Groin Injury

Quikclot has been used many years for treating external hemorrhage in the battle field. In this study, the hemostatic performance of NZG-JY (natural zeolite granules from Jinyun, China) was evaluated and compared with Quikclot in a lethal rabbit model of complex groin injury. Fifty-six anesthetized rabbits were randomized to three different groups: (1) NZG-JY (n = 19); (2) Quikclot (n = 19); an...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • The Journal of surgical research

دوره 169 1  شماره 

صفحات  -

تاریخ انتشار 2011