Dosimetric evaluation of radiation dose rate effect in respiratory gated intensity modulated radiation therapy†

نویسندگان

  • C Khamfongkhruea
  • C Tannanonta
  • S Thongsawad
چکیده

PURPOSE To investigate the dosimetric accuracy of the sliding window gated IMRT compared with the static treatment, using varying dose rates. MATERIALS AND METHODS This study measured changes in output and diode array response with changing dose rate, verified the precision of the motion table, and measured changes in dose distribution accuracy with film and diodes at two depths with changing dose rate. During 4DCT (4 Dimensional Computed Tomography), the patient's respiratory signals and target motion were recorded and imported to the XY4D simulation table of SUN NUCLEAR Corporation to simulate the patient's respiration and tumour motion. A single field of each sliding window IMRT plan with 30º wedge and one for lung cancer were used in this study. Three irradiating conditions, static and moving target with and without gating, were applied to both plans. RESULTS The standard deviations of output, with the dose rates changing from 300-600 MU/min, were 0.065 cGy and 0.169 cGy for the ionisation chamber and diode, respectively. The verification of the motion table shows very good precision with 9.98 ± 0.02 cm (true value = 10.0 cm). The measurements by MapCheck show the gamma index of the planned absolute dose distribution in static and moving targets with gating, resulting in more than 96% passing for all dose rates. The absolute dose distribution measured by film for the static target was agreeable with the value of moving target with gating. CONCLUSION The sliding window gated IMRT technique is able to deliver an accurate dose to a moving target with the dose rate of 300-600 MU/min that is suitable for clinical treatment.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Investigation of lung normal tissue doses in lung tumors radiation therapy using both gated and conventional radiotherapy

Introduction: In radiation therapy of lung tumors, respiratory motion causes target moving, so a larger margin is needed to cover the clinical target volume (CTV). With the margin increasing, a larger volume of normal tissue will be exposed to high-dose.  In this study, dosimetric parameters of normal lung tissue were compared between gated and conventional radiotherapy (RT), u...

متن کامل

The Dosimetric Effects of Different Multileaf Collimator Widths on Physical Dose Distributions

Introduction: Geometric changes in the multileaf collimator (MLC) led to dosimetric considerations in intensity-modulated radiation therapy (IMRT) due to the number and size of the pixels in the intensity map, which are determined by the MLC leaf width. In this study, we evaluated the dosimetric effects of different MLC widths on physical dose distributions for IMRT plans. Materials and Method...

متن کامل

Respiratory motion effect on tumor and normal tissue doses in patients with lung cancer, treated with Intensity Modulation Radiation Therapy and Three Dimensional Conformal Radiation Therapy.

Introduction:   The aim of this study is to investigate the effect of respiratory motion during radiation therapy in patient with lung cancer and comparison of dosimetric parameters between Intensity modulation radiation therapy and three-dimensional conformal radiotherapy in lung cancer.   Materials and Methods:   Two CT scan was performred for each pati...

متن کامل

Dosimetric evaluations of the interplay effect in respiratory-gated intensity-modulated radiation therapy.

The interplay between a mobile target and a dynamic multileaf collimator can compromise the accuracy of intensity-modulated radiation therapy (IMRT). Our goal in this study is to investigate the dosimetric effects caused by the respiratory motion during IMRT. A moving phantom was built to simulate the typical breathing motion. Different sizes of the gating windows were selected for gated delive...

متن کامل

The evaluation of lung doses for radiation pneumonia risk in stereotactic body radiotherapy: A comparison of intensity modulated radiotherapy, intensity modulated arc therapy, cyberknife and helical tomotherapy

Background: Radiation Pneumonia (RP) is one of the most extensive side effects in Stereotactic Body Radiotherapy (SBRT) of lung cancer. SBRT are performed by means of Intensity Modulated Radiotherapy (IMRT), Intensity Modulated Arc Therapy (IMAT), CyberKnife (CK) or Helical Tomotherapy (HT) treatment methods. In this study, we performed a plan study to determine the plan parameter such as the M...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 8  شماره 

صفحات  -

تاریخ انتشار 2012