Dosimetry limitations and pre-treatment dose profile correction for sliding window IMRT

نویسندگان

  • G. Grigorov
  • J.C.L. Chow
  • N. Yazdani
چکیده

*Corresponding author: Dr. Grigor Grigorov, Medical Physics Department, Grand River Regional, Cancer Center, PO Box 9056, 835 King St West, Kitchener, ON, Canada N2G 1G3. E-mail: [email protected] Background: This work investigated the dosimetry limitations of the random and systematic uncertainties of sliding window (SW) intensity modulated radiation therapy (IMRT). Materials and Methods: A Varian 21EX linear accelerator, Pinnacle3 treatment planning system and radiographic film dosimetry was used. The limitations of the SW were studied using beam modulation ranging from 2 to 100 MU/beam, DR from 100 to 600 MU min-1, LV from 1 to 5 cm s-1 and field size up to 12 × 12 cm2. The random and systematic errors were investigated using clinical and flat beams, as well as beams of high profile modulation including linear, exponential, and sinusoidal profiles. Results: The leading edge and plateau of the SW profiles have a significant deformation for higher DR and for beams of < 10 MUs/beam. It was found that the error is directly proportional to the DR and LV, and inversely proportional to the number of MU/beam. Conclusion: The high DR and LV are limiting factors, producing random profile deformation when SW beams of small number of MU/beam are delivered. A very good agreement was found between the planned and delivered geometrical and clinical dose profiles when beams > 10 MUs irradiated by a DR from 100 to 600 MU min-1 and LV from 1 to 5 cm s-1. After the proposed correction, an average difference < 0.5% for clinical profiles was measured for beams irradiated with DR = 600 MU min-1 and LV= 5 cm s-1. It was concluded that this correction methodology may serve as a pre-treatment Quality Assurance tool for SW IMRT beams. Iran. J. Radiat. Res., 2010; 8 (2): 61­74

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Dosimetry limitations and pre-treatment dose profile correction for sliding window IMRT

Background: This work investigated the dosimetry limitations of the random and systematic uncertainties of sliding window (SW) intensity modulated radiation therapy (IMRT). Materials and Methods: A Varian 21EX linear accelerator, Pinnacle3 treatment planning system and radiographic film dosimetry was used. The limitations of the SW were studied using beam modulation ranging from 2 to 10...

متن کامل

Dosimetric evaluation of field-in-field and sliding-window IMRT in endometrium cancer patients with a new approach for the conformity index

Background: Purpose of the study is to evaluate field-in-field (FIF) and sliding-window intensity modulated radiation therapy (SW-IMRT) treatment plans in terms of homogeneity index (HI), conformity index (CI) and normal tissue doses for organs at risk (OAR) in patients with endometrial carcinoma along with deriving a new expression for CI. Materials and Methods: Four different FIF plans with d...

متن کامل

Evaluation of a fast method of EPID-based dosimetry for IMRT and Comparison with 3D EPID-based dosimetry system using conventional two- and three-dimensional detectors for VMAT

Introduction: Electronic portal imaging devices (EPIDs) could potentially be useful for intensity-modulated radiation therapy (IMRT) and VMAT QA. The data density, high resolution, large active area, and efficiency of the MV EPID make it an attractive option. However, EPIDs were designed as imaging devices, not dosimeters, and as a result they do not inherently measure dose in ...

متن کامل

The effect of the target-organ geometric complexity on the choice of delivery between RapidArc and sliding-window IMRT for nasopharyngeal carcinoma.

We attempted to assess the effect of target-organ geometric complexity on the plan quality of sliding-window intensity-modulated radiotherapy (IMRT), double-arc (RA2), and triple-arc (RA3) RapidArc volumetric-modulated arc radiotherapy for nasopharyngeal carcinoma (NPC). Plans for 9-field sliding-window IMRT, RA2, and RA3 were optimized for 36 patients with NPC ranging from T1 to T4 tumors. Ini...

متن کامل

The Quality Control of Intensity Modulated Radiation Therapy (IMRT) for ONCOR Siemens Linear Accelerators Using Film Dosimetry

Introduction Intensity Modulated Radiation Therapy (IMRT) has made a significant progress in radiation therapy centers in recent years. In this method, each radiation beam is divided into many subfields that create a field with a modulated intensity. Considering the complexity of this method, the quality control for IMRT is a topic of interest for researchers. This article is about the various ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -