Experimental evaluation of the Medtronic Maxima Forté hollow fiber membrane oxygenator.

نویسندگان

  • X M Mueller
  • H T Tevaearai
  • M Augstburger
  • J Horisberger
  • L K von Segesser
چکیده

A new hollow fiber membrane oxygenator, the Medtronic Maxima Forté, was tested for gas transfer, blood path resistance and blood handling characteristics in a standardized setting with surviving animals. Three calves (mean body weight: 71 +/- 9.6 kg) were placed on cardiopulmonary bypass at a mean flow rate of 50 ml/kg/min for six hours. The circuit included the Maxima Forté oxygenator. The animals were weaned from cardiopulmonary bypass and then from the ventilator. After seven days, the animals were sacrificed electively. Physiologic blood gas values could be maintained throughout perfusion in all animals. Mean pressure drop through the oxygenator varied between 49 mmHg and 66 mmHg. The respective baseline values for red blood cell count, white blood cell count and platelets were 8.90 +/- 1.26 10(6)/mm3, 7.46 +/- 3.17 10(3)/mm3. and 680 +/- 216 10(3)/mm3. Red blood cell and platelet counts dropped slightly to 7.26 +/- 1.61 10(6)/mm3 and 400 +/- 126 10(3)/mm3 at the end of the bypass, whereas the white blood cell count increased up to 9.13 +/- 5.25 10(3)/mm3. All three cell lines returned to near their baseline values after seven days. Blood trauma evaluated as a function of plasma hemoglobin (plasma Hb) and lactate dehydrogenase (LDH) showed stable values during all the perfusion time. Both peaked at 24 hours before returning to their baseline values at seven days. LDH showed a statistically significant variation: 3255 +/- 693 IU at 24 hours versus 2029 +/- 287 IU at baseline (p = 0.04). The variation of plasma Hb was not statistically significant (93.5 +/- 7.7 mumol/l at 24 hours versus 77.3 +/- 52.3 mumol/l at baseline) indicating a weak effect of the perfusion on blood trauma. The Medtronic Maxima Forté hollow fiber membrane oxygenator offered good gas exchange capabilities, a low pressure drop, and low blood trauma over a prolonged perfusion time of six hours in this evaluation.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Retrospective analysis comparing hollow fiber and silicone membrane oxygenators for neonates on ECMO.

There is little information showing the use of microporous polypropylene hollow fiber oxygenators during extra-corporeal life support (ECLS). Recent surveys have shown increasing use of these hollow fibers amongst ECLS centers in the United States. We performed a retrospective analysis comparing the Terumo BabyRx hollow fiber oxygenator to the Medtronic 800 silicone membrane oxygenator on 14 ne...

متن کامل

Development of a new hollow fiber silicone membrane oxygenator for ECMO: the recent progress.

Throughout the last 50 years, many improvements have been made for a more effective oxygenator. A large plate type membrane oxygenator, used by Clowes, and a coil type, used by Kolff, has evolved into the small hollow fiber oxygenator. The complex bubble oxygenator, or rotating disk oxygenator, has become a small disposable bubble oxygenator. The currently available oxygenators are extremely sm...

متن کامل

A clinical evaluation of the Terumo Capiox SX18R hollow fiber oxygenator.

The Terumo Capiox SX18R is a commercially available, low prime, reverse phase, hollow fiber membrane oxygenator. The oxygenator consists of a 1.8 m2 microporous polypropylene hollow fiber bundle, a 2200 cm2 tubular stainless steel heat exchanger, and an open hard shell venous reservoir with integral cardiotomy filter. The Terumo Capiox SX18R oxygenator was evaluated to determine its clinical ox...

متن کامل

Preclinical evaluation of a new hollow fiber silicone membrane oxygenator for pediatric cardiopulmonary bypass: ex-vivo study.

Based on the results of many experimental models, a hollow fiber silicone membrane oxygenator applicable for long-term extracorporeal membrane oxygenation (ECMO) was developed. For further high performance and antithrombogenicity, this preclinical model was modified, and a new improved oxygenator was successfully developed. In addition to ECMO application, the superior biocompatibility of silic...

متن کامل

Gas transfer performance of a hollow fiber silicone membrane oxygenator: ex vivo study.

Based on the results of in vitro studies of many experimental models, a silicone hollow fiber membrane oxygenator for pediatric cardiopulmonary bypass (CPB) and extracorporeal membrane oxygenation (ECMO) was developed using an ultrathin silicone hollow fiber with a 300 microm outer diameter and a wall thickness of 50 microm. In this study, we evaluated the gas transfer performance of this oxyge...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • The journal of extra-corporeal technology

دوره 30 3  شماره 

صفحات  -

تاریخ انتشار 1998