Multiplex RARE Dixon: a novel multislice RARE sequence applied to simultaneous slice fat-water Dixon imaging
نویسندگان
چکیده
Introduction Multislice excitation methods have been widely used to increase efficiency in volume coverage. In the simultaneous spin-echo refocusing sequence (SER) [1], two or more slices are excited simultaneously and refocused with a train of 180° pulses which act simultaneously on both slices. However, the SER sequence does not fulfil CPMG conditions, and requires special spoiler gradients to eliminate undesired stimulated components. This leads to signal loss and limits echo train length. Recently, a modification of SER was introduced, called multiplex RARE (mRARE) [2], which allows two spin-echo trains to be simultaneously excited and interleaved, while satisfying CPMG conditions. Here we show an application of mRARE in a novel, fast Dixon fat-water sequence which acquires two slices simultaneously. The sequence is similar to the GRASE-Dixon sequence reported in Ref [3], but although Ref [3] fulfils the CPMG condition for timing, it does not fulfil the CPMG condition for phase accumulation and shows artifacts from stimulated echoes. Technique The mRARE Dixon sequence is shown in Fig. 1. Two slices are simultaneously excited and refocused by an interleaved train of their own slice selective pulses. At the spin-echo for one slice, e.g. S1, only the signal from slice 1 has been fully rephased while slice 2 remains dephased in the slice and phase directions. Rephasing slice 2 will dephase slice 1 and vice versa. There will be crosstalk between slices if there is insufficient dephasing because of susceptibility gradients or if unwanted signal from imperfect refocusing pulses are insufficiently crushed by spoiler gradients A and B. The spin-echoes occur halfway between their refocusing pulses, and by considering the gradient moments, it can be seen that the CPMG condition for phase accumulation is also satisfied. Thus stimulated echoes will coincide with the spin-echoes as with RARE. Note that Dixon images with symmetric sampling have some redundancy [4], and that out-of-phase gradient echoes need only be collected on one side of the spin-echo. Therefore, it is possible to fit in echoes from another slice on the other side of the spin-echo. In Fig. 1, a gradientand a spin-echo from different slices are collected on one side of each of the spin-echoes. Two-point Dixon fat-water separation can then be performed.
منابع مشابه
Faster fat-water imaging with a novel multislice time-shifted GRASE-Dixon sequence
Introduction Recently, the gradientand spin-echo (GRASE) technique has been used to increase efficiency in Dixon fat-water imaging [1,2]. In these sequences, oscillating gradient lobes are carefully positioned between spin refocusing pulses to acquire multiple echoes with the necessary fat-water phase shifts. As it is necessary to collect at least two echoes with approximately π fat-water phase...
متن کاملBreast MRI Using a Balanced Steady-State Free Precession Imaging with a Two-Point Dixon Fat-Water Reconstruction Algorithm: Preliminary Experience and Comparison with 2D FSE
INTRODUCTION: Balanced steady-state free precession (b-SSFP) techniques yielding high SNR images in short scan times with T2-like image contrast have several appealing attributes for clinical breast MRI [1]. Clinical T2 imaging in the breast is routinely performed using a fat suppressed 2D Fast Spin Echo (FSE) pulse sequence with high in-plane spatial resolution. The 3-5 mm slice thickness obsc...
متن کاملFast spin-echo triple-echo dixon (fTED) technique for efficient T2-weighted water and fat imaging.
Previously published fast spin-echo (FSE) implementations of a Dixon method for water and fat separation all require multiple scans and thus a relatively long scan time. Further, the minimum echo spacing (esp), a time critical for FSE image quality and scan efficiency, often needs to be increased in order to bring about the required phase shift between the water and fat signals. This work propo...
متن کاملDistortion correction of body diffusion weighted images using three point-Dixon method
Introduction: Significant distortion can often be observed in body diffusion weighted images (DWI) using single shot EPI. Distortion is caused by external magnetic field (B0) inhomogeneity. The degree of distortion is proportional to B0. When a precise B0 map is obtained, distortion can be corrected using the B0 map. In this study, we demonstrate that a B0 map obtained using three point-Dixon m...
متن کاملThree images from two echoes: Reconstruction of water, fat and silicone images using a combined one-point and two-point Dixon approach. Application to 3D GRE in breast implant imaging
Introduction: MR imaging of breast implants has emerged as an important tool basically due to two reasons a) the difficulty to diagnose lesions in augmented or reconstructed breasts solely using mammography and ultrasound b) the necessity to assess the intactness of the implant on a regular basis [1]. A conventional complete exam of a patient with silicone implants to detect possible ruptures c...