Acute posttraumatic catastrophic failure of a second-generation, highly cross-linked ultra-high-molecular-weight polyethylene patellar component.

نویسندگان

  • Matthew J Goldstein
  • Michael P Ast
  • Frank R Dimaio
چکیده

A 74-year-old man underwent total knee arthroplasty in 1996 and subsequent revision total knee arthroplasty for aseptic loosening in 2005. During revision, an all-polyethylene, 7-mm, second-generation, highly cross-linked ultra-high-molecular-weight polyethylene patellar component (Durasul; Zimmer, Warsaw, Indiana) was used. The patient recovered well, but he presented with severe acute pain after a popping feeling was detected during a game of golf in postoperative year 4. Radiographs demonstrated a fracture of the patellar component. At re-revision surgery, polyethylene fragments were encountered without visual evidence of wear. Appropriate rotational and axial alignment of the components was confirmed. Patellar revision was performed. To the authors' knowledge, this is the first report of an acute posttraumatic catastrophic fracture of a second-generation, highly cross-linked ultra-high-molecular-weight polyethylene patellar component. Although no malposition of the components was noted, abnormal tensile forces across the patellar component can be transmitted by altered patellofemoral kinematics. This combination could lead to a small surface crack that propagates deeper into the component with continued impaction and an eventual acute fracture of the entire component. The patellar component failed in a piecemeal fashion following acute trauma. Although many advantages exist to the enhancement of polyethylene patellar components, including the potential for improved wear characteristics, cross-linking has reduced resistance to crack initiation and propagation. Therefore, fracture risk should be considered when using second-generation, highly cross-linked ultra-high-molecular-weight polyethylene.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Radiostereometric analysis study of tantalum compared with titanium acetabular cups and highly cross-linked compared with conventional liners in young patients undergoing total hip replacement.

BACKGROUND Radiostereometric analysis provides highly precise measurements of component micromotion relative to the bone that is otherwise undetectable by routine radiographs. This study compared, at a minimum of five years following surgery, the micromotion of tantalum and titanium acetabular cups and femoral head penetration in highly cross-linked polyethylene liners and conventional (ultra-h...

متن کامل

Basic kinematics and biomechanics of the patellofemoral joint part 2: the patella in total knee arthroplasty.

Patellar and femoral component in total knee arthroplasty are inextricably linked as a functional unit. The configuration of this unit has been a matter of ongoing debate, and the myriad of different patellar and femoral components currently available reflect the lack of consensus with respect to the ideal design. One of the major challenges is to overcome the biomechanical disadvantages of a s...

متن کامل

Delamination of a highly cross-linked polyethylene liner associated with titanium deposits on the cobalt-chromium modular femoral head following dislocation.

Retrieval studies of total hip replacements with highly cross-linked ultra-high-molecular-weight polyethylene liners have shown much less surface damage than with conventional ultra-high-molecular-weight polyethylene liners. A recent revision hip replacement for recurrent dislocation undertaken after only five months revealed a highly cross-linked polyethylene liner with a large area of visible...

متن کامل

Catastrophic failure of the Elite Plus total hip replacement, with a Hylamer acetabulum and Zirconia ceramic femoral head.

We report catastrophic early failure of a cemented total hip replacement comprising a modular femoral component with a Zirconia ceramic head and an acetabular component of cross-linked ultra-high molecular-weight polyethylene (Hylamer). Between 1995 and 1999 we implanted 29 hips in 26 patients with a mean age of 49.2 years. Survivorship analysis in this group revealed a failure rate of 67.6% at...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Orthopedics

دوره 35 7  شماره 

صفحات  -

تاریخ انتشار 2012