Comparison of Image Quality of Shoulder CT Arthrography Conducted Using 120 kVp and 140 kVp Protocols
نویسندگان
چکیده
OBJECTIVE To compare the image quality of shoulder CT arthrography performed using 120 kVp and 140 kVp protocols. MATERIALS AND METHODS Fifty-four CT examinations were prospectively included. CT scans were performed on each patient at 120 kVp and 140 kVp; other scanning parameters were kept constant. Image qualities were qualitatively and quantitatively compared with respect to noise, contrast, and diagnostic acceptability. Diagnostic acceptabilities were graded using a one to five scale as follows: 1, suboptimal; 2, below average; 3, acceptable; 4, above average; and 5, superior. Radiation doses were also compared. RESULTS Contrast was better at 120 kVp, but noise was greater. No significant differences were observed between the 120 kVp and 140 kVp protocols in terms of diagnostic acceptability, signal-to-noise ratio, or contrast-to-noise ratio. Lowering tube voltage from 140 kVp to 120 kVp reduced the radiation dose by 33%. CONCLUSION The use of 120 kVp during shoulder CT arthrography reduces radiation dose versus 140 kVp without significant loss of image quality.
منابع مشابه
Cardiac helical CT involving a low-radiation-dose protocol with a 100-kVp setting
To compare the radiation dose and image quality of retrospective electrocardiogram (ECG)-gated cardiac computed tomography (CT) between a 100-kVp protocol, hybrid iterative reconstruction (HIR), and display preset optimization and the 120-kVp protocol.We prospectively enrolled 100 patients with tachycardia or atrial fibrillation scanned retrospective ECG-gated cardiac CT. We randomly assigned 5...
متن کاملLow kilovoltage CT of the neck with 70 kVp: comparison with a standard protocol.
BACKGROUND AND PURPOSE CT protocols should aim for radiation doses being as low as reasonably achievable. The purpose of our study was to assess the image quality and radiation dose of neck CT at a tube potential of 70 kVp. MATERIALS AND METHODS Twenty patients (7 female, mean age 51.4 years, age range 19-81 years) underwent contrast-enhanced 64-section CT of the neck at 70 kVp (ATCM, effecti...
متن کامل80-kVp CT Using Iterative Reconstruction in Image Space Algorithm for the Detection of Hypervascular Hepatocellular Carcinoma: Phantom and Initial Clinical Experience
OBJECTIVE To investigate whether the low-tube-voltage (80-kVp), intermediate-tube-current (340-mAs) MDCT using the Iterative Reconstruction in Image Space (IRIS) algorithm improves lesion-to-liver contrast at reduced radiation dosage while maintaining acceptable image noise in the detection of hepatocellular carcinomas (HCC) in thin (mean body mass index, 24 ± 0.4 kg/m(2)) adults. SUBJECTS AN...
متن کاملThe Image Quality and Radiation Dose of 100-kVp versus 120-kVp ECG-Gated 16-Slice CT Coronary Angiography
OBJECTIVE This study was conducted to assess the feasibility of performing 100-kVp electrocardiogram (ECG)-gated coronary CT angiography, as compared to 120-kVp ECG-gated coronary CT angiography. MATERIALS AND METHODS We retrospectively evaluated one hundred eighty five gender- and body mass index-matched 16-slice coronary CT sets of data, which were obtained using either 100 kVp and 620 effe...