The clinical effectiveness and cost-effectiveness of point-of-care tests (CoaguChek system, INRatio2 PT/INR monitor and ProTime Microcoagulation system) for the self-monitoring of the coagulation status of people receiving long-term vitamin K antagonist therapy, compared with standard UK practice: systematic review and economic evaluation.

نویسندگان

  • Pawana Sharma
  • Graham Scotland
  • Moira Cruickshank
  • Emma Tassie
  • Cynthia Fraser
  • Chris Burton
  • Bernard Croal
  • Craig R Ramsay
  • Miriam Brazzelli
چکیده

BACKGROUND Self-monitoring (self-testing and self-management) could be a valid option for oral anticoagulation therapy monitoring in the NHS, but current evidence on its clinical effectiveness or cost-effectiveness is limited. OBJECTIVES We investigated the clinical effectiveness and cost-effectiveness of point-of-care coagulometers for the self-monitoring of coagulation status in people receiving long-term vitamin K antagonist therapy, compared with standard clinic monitoring. DATA SOURCES We searched major electronic databases (e.g. MEDLINE, MEDLINE In Process & Other Non-Indexed Citations, EMBASE, Bioscience Information Service, Science Citation Index and Cochrane Central Register of Controlled Trials) from 2007 to May 2013. Reports published before 2007 were identified from the existing Cochrane review (major databases searched from inception to 2007). The economic model parameters were derived from the clinical effectiveness review, other relevant reviews, routine sources of cost data and clinical experts' advice. REVIEW METHODS We assessed randomised controlled trials (RCTs) evaluating self-monitoring in people with atrial fibrillation or heart valve disease requiring long-term anticoagulation therapy. CoaguChek(®) XS and S models (Roche Diagnostics, Basel, Switzerland), INRatio2(®) PT/INR monitor (Alere Inc., San Diego, CA USA), and ProTime Microcoagulation system(®) (International Technidyne Corporation, Nexus Dx, Edison, NJ, USA) coagulometers were compared with standard monitoring. Where possible, we combined data from included trials using standard inverse variance methods. Risk of bias assessment was performed using the Cochrane risk of bias tool. A de novo economic model was developed to assess the cost-effectiveness over a 10-year period. RESULTS We identified 26 RCTs (published in 45 papers) with a total of 8763 participants. CoaguChek was used in 85% of the trials. Primary analyses were based on data from 21 out of 26 trials. Only four trials were at low risk of bias. Major clinical events: self-monitoring was significantly better than standard monitoring in preventing thromboembolic events [relative risk (RR) 0.58, 95% confidence interval (CI) 0.40 to 0.84; p = 0.004]. In people with artificial heart valves (AHVs), self-monitoring almost halved the risk of thromboembolic events (RR 0.56, 95% CI 0.38 to 0.82; p = 0.003) and all-cause mortality (RR 0.54, 95% CI 0.32 to 0.92; p = 0.02). There was greater reduction in thromboembolic events and all-cause mortality through self-management but not through self-testing. Intermediate outcomes: self-testing, but not self-management, showed a modest but significantly higher percentage of time in therapeutic range, compared with standard care (weighted mean difference 4.44, 95% CI 1.71 to 7.18; p = 0.02). Patient-reported outcomes: improvements in patients' quality of life related to self-monitoring were observed in six out of nine trials. High preference rates were reported for self-monitoring (77% to 98% in four trials). Net health and social care costs over 10 years were £7295 (self-monitoring with INRatio2); £7324 (standard care monitoring); £7333 (self-monitoring with CoaguChek XS) and £8609 (self-monitoring with ProTime). The estimated quality-adjusted life-year (QALY) gain associated with self-monitoring was 0.03. Self-monitoring with INRatio2 or CoaguChek XS was found to have ≈ 80% chance of being cost-effective, compared with standard monitoring at a willingness-to-pay threshold of £20,000 per QALY gained. CONCLUSIONS Compared with standard monitoring, self-monitoring appears to be safe and effective, especially for people with AHVs. Self-monitoring, and in particular self-management, of anticoagulation status appeared cost-effective when pooled estimates of clinical effectiveness were applied. However, if self-monitoring does not result in significant reductions in thromboembolic events, it is unlikely to be cost-effective, based on a comparison of annual monitoring costs alone. Trials investigating the longer-term outcomes of self-management are needed, as well as direct comparisons of the various point-of-care coagulometers. STUDY REGISTRATION This study is registered as PROSPERO CRD42013004944. FUNDING The National Institute for Health Research Health Technology Assessment programme.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Is self-monitoring an effective option for people receiving long-term vitamin K antagonist therapy? A systematic review and economic evaluation

OBJECTIVES To investigate the clinical and cost-effectiveness of self-monitoring of coagulation status in people receiving long-term vitamin K antagonist therapy compared with standard clinic care. DESIGN Systematic review of current evidence and economic modelling. DATA SOURCES Major electronic databases were searched up to May 2013. The economic model parameters were derived from the clin...

متن کامل

First PT/INR home self-testing device with Bluetooth technology enabling remote care programmes for anticoagulated patients

Roche (SIX: RO, ROG; OTCQX: RHHBY) announced today that it has launched the CoaguChek® INRange system in countries accepting the CE Mark. The CoaguChek INRange system is the first Bluetooth enabled PT/INR home health device that helps patients and their healthcare providers (HCPs) have greater control over their coagulation status and the ability to monitor Vitamin K Antagonist (VKA) therapy. P...

متن کامل

The Clinical Impact of Different Coagulometers on Patient Outcomes

INTRODUCTION Long-term anticoagulation therapy using vitamin K antagonists (VKA) is used in millions of patients worldwide to reduce the risk of thrombotic or thromboembolic events. Control and monitoring of VKA therapy is improved by the regular self-measurement of international normalized ratio (INR) using a home monitoring device. This retrospective analysis of a large cohort of patients in ...

متن کامل

The Effectiveness of a Chronic Disease Self-Management Program for Elderly People: a Systematic Review

Introduction: Elderly population is increasing in Pakistan. Majority of people aged 50 – 64, suffer from two or more chronic conditions. Care of elderly people is sub-optimal. Healthcare system has limited resources to provide healthcare support to elderly people. Thus, there is need to empower the elderly to self-manage their health conditions. Chronic Disease Self-Management Program (CDSMP) i...

متن کامل

Accuracy of the CoaguChek XS for point-of-care international normalized ratio (INR) measurement in children requiring warfarin.

Point-of-care INR (POC INR) meters can provide a safe and effective method for monitoring oral vitamin K antagonists (VKAs) in children. Stollery Children's Hospital has a large POC INR meter loan program for children requiring oral VKAs. Our protocol requires that POC INR results be compared to the standard laboratory INR for each child on several consecutive tests to ensure accuracy of CoaguC...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Health technology assessment

دوره 19 48  شماره 

صفحات  -

تاریخ انتشار 2015