Calculation of the index of microcirculatory resistance without coronary wedge pressure measurement in the presence of epicardial stenosis.

نویسندگان

  • Andy S Yong
  • Jamie Layland
  • William F Fearon
  • Michael Ho
  • Maulik G Shah
  • David Daniels
  • Robert Whitbourn
  • Andrew Macisaac
  • Leonard Kritharides
  • Andrew Wilson
  • Martin K Ng
چکیده

OBJECTIVES This study sought to investigate a novel method to calculate the index of microcirculatory resistance (IMR) in the presence of significant epicardial stenosis without the need for balloon dilation to measure the coronary wedge pressure (P(w)). BACKGROUND The IMR provides a quantitative measure of coronary microvasculature status. However, in the presence of significant epicardial stenosis, IMR calculation requires incorporation of the coronary fractional flow reserve (FFR(cor)), which requires balloon dilation within the coronary artery for P(w) measurement. METHODS A method to calculate IMR by estimating FFR(cor) from myocardial FFR (FFR(myo)), which does not require P(w) measurement, was developed from a derivation cohort of 50 patients from a single institution. This method to calculate IMR was then validated in a cohort of 72 patients from 2 other different institutions. Physiology measurements were obtained with a pressure-temperature sensor wire before coronary intervention in both cohorts. RESULTS From the derivation cohort, a strong linear relationship was found between FFR(cor) and FFR(myo) (FFR(cor) = 1.34 × FFR(myo) - 0.32, r(2) = 0.87, p < 0.001) by regression analysis. With this equation to estimate FFR(cor) in the validation cohort, there was no significant difference between IMR calculated from estimated FFR(cor) and measured FFR(cor) (21.2 ± 12.9 U vs. 20.4 ± 13.6 U, p = 0.161). There was good correlation (r = 0.93, p < 0.001) and agreement by Bland-Altman analysis between calculated and measured IMR. CONCLUSIONS The FFR(cor), and, by extension, microcirculatory resistance can be derived without the need for P(w). This method enables assessment of coronary microcirculatory status before or without balloon inflation, in the presence of epicardial stenosis.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Coronary microcirculatory resistance is independent of epicardial stenosis.

BACKGROUND Recent studies show that coronary microcirculatory impairment is an independent predictor of poor outcomes in patients with cardiovascular disease. However, controversy exists over whether microcirculatory resistance, a measure of coronary microcirculatory status, is dependent on epicardial stenosis severity. Previous studies demonstrating that microcirculatory resistance is dependen...

متن کامل

Microvascular resistance is not influenced by epicardial coronary artery stenosis severity: experimental validation.

BACKGROUND The effect of epicardial artery stenosis on myocardial microvascular resistance remains controversial. Recruitable collateral flow, which may affect resistance, was not incorporated into previous measurements. METHODS AND RESULTS In an open-chest pig model, distal coronary pressure was measured with a pressure wire, and the apparent minimal microvascular resistance was calculated d...

متن کامل

Epicardial stenosis severity does not affect minimal microcirculatory resistance.

BACKGROUND Whether minimal microvascular resistance of the myocardium is affected by the presence of an epicardial stenosis is controversial. Recently, an index of microcirculatory resistance (IMR) was developed that is based on combined measurements of distal coronary pressure and thermodilution-derived mean transit time. In normal coronary arteries, IMR correlates well with true microvascular...

متن کامل

Novel index for invasively assessing the coronary microcirculation.

BACKGROUND A relatively simple, invasive method for quantitatively assessing the status of the coronary microcirculation independent of the epicardial artery is lacking. METHODS AND RESULTS By using a coronary pressure wire and modified software, it is possible to calculate the mean transit time of room-temperature saline injected down a coronary artery. The inverse of the hyperemic mean tran...

متن کامل

Magnetic Resonance for Noninvasive Detection of Microcirculatory Disease Associated With Allograft Vasculopathy: Intracoronary Measurement Validation.

INTRODUCTION AND OBJECTIVES Cardiac allograft vasculopathy affects both epicardial and microcirculatory coronary compartments. Magnetic resonance perfusion imaging has been proposed as a useful tool to assess microcirculation mostly outside the heart transplantation setting. Instantaneous hyperemic diastolic flow velocity-pressure slope, an intracoronary physiology index, has demonstrated a bet...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • JACC. Cardiovascular interventions

دوره 6 1  شماره 

صفحات  -

تاریخ انتشار 2013