SU-E-I-29: Dose Rate and Image Quality Comparison Between Philips Veradius and GE OEC 9900 Elite Fluoroscopic C-Arms.

نویسندگان

  • A Springer
  • E Solem
  • R Kobistek
چکیده

PURPOSE The radiation dose rate and image quality were compared on GE OEC 9900 Elite Vas 8 and Philips Veradius fluoroscopic C-arms. METHODS Plastic (PMMA) phantoms were used to simulate various thicknesses of soft tissues encountered during fluoroscopically-guided interventional procedures while aluminum and copper phantoms were used to simulate extremities of different sizes. An image quality test tool and an ionization chamber were included in the setup of each phantom. The phantoms were imaged in several fluoroscopy modes and at different dose levels. The entrance surface air KERMA rate (ESAKR) was measured for each phantom in each fluoroscopy mode. Simultaneously, the image quality was assessed by counting the number of low-contrast objects visible in the image quality test tool. For a given phantom type and thickness, the ESAKR was plotted versus the number of visible low-contrast objects to compare the dose rate required for a given level of image quality. Additionally, the mean ESAKR per number of objects was calculated and p-values were used to test for statistically significant differences between the C-arms. RESULTS For comparable image quality with a given phantom type and thickness the ESAKR with the OEC was typically much higher than with the Veradius. One-tailed, unpaired t-tests confirm that the mean ESAKR per number of visible low-contrast objects was lower for the Veradius than for the OEC with all thicknesses of PMMA (p < 0.05). The differences for the aluminum and copper phantoms were much smaller and were not statistically significant (p-values ranged from 0.2 to 0.47). CONCLUSION In general, the Veradius C-arm provided better low-contrast image quality during fluoroscopy, with lower ESAKR than the OEC 9900 Elite Vas 8. The differences were statistically significant only with PMMA phantoms. Funding support provided by Philips Healthcare (Bothell, WA).

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Accuracy of C-arm measurements in assessment of paediatric femoral fracture shortening

Purpose Spica casting of femoral shaft fractures is standard practice for young children. In 1983, Dr. Rang published an illustration demonstrating that radiographs can be misleading when measuring femoral fracture shortening. Today, fluoroscopy is used to determine shortening after closed reduction and spica casting. This paper aims to determine the accuracy of fluoroscopy when measuring 25 mm...

متن کامل

Functionality and operation of fluoroscopic automatic brightness control/automatic dose rate control logic in modern cardiovascular and interventional angiography systems: a report of Task Group 125 Radiography/Fluoroscopy Subcommittee, Imaging Physics Committee, Science Council.

Task Group 125 (TG 125) was charged with investigating the functionality of fluoroscopic automatic dose rate and image quality control logic in modern angiographic systems, paying specific attention to the spectral shaping filters and variations in the selected radiologic imaging parameters. The task group was also charged with describing the operational aspects of the imaging equipment for the...

متن کامل

Amitraz Poisoning; A case study

A m i t r a z, a n i ns e c t i c i d e /a ca ri c i de of the f o r m a m i d i n e p e st i c i d e s group, is a ? 2 a d r e n e r g i c ag on i st a nd of t he a m i d i ne c h e m i ca l f a m il y generally us e d to c o n t r ol animal e c top a r a s i t e s. Poisoning due to am i t r a z i s r a r e and character...

متن کامل

Assessment the effect of CT localizer parameter on radiation dose and image quality of GE CT spiral images, A phantom study

Introduction: Using computed tomography (CT) is increasing significantly in Iran and worldwide. Reduction of CT's radiation dose is a critical concern. The aim of this study was evaluating the effect of localizer kVp on spiral CT images when the tube current modulation (TCM) is activated. Materials and Methods: A special wedding cake phantom with circular in...

متن کامل

Low kilovolt “prospective ECG-triggering” vs. “retrospective ECG-gating” coronary CTA: comparison of image quality and radiation dose

Background: To compare image quality and radiation doses of low kilovolt (kV) “prospective ECG-triggering” (PT) and standard “retrospective ECG-gating” (RG) coronary computed tomography (CT) angiography. Materials and Methods: A total of 101 consecutive patients (76 males, 25 females; mean age: 55.44 ± 8.28 years) with low-to-intermediate risk status for coronary artery disease and with a body ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Medical physics

دوره 39 6Part4  شماره 

صفحات  -

تاریخ انتشار 2012