Fluoroscopic Exposure With Use of Mini-C-Arm During Routine Hand Surgery: A Prospective Comparison of Hand Versus Eye Radiation Dosage.
نویسندگان
چکیده
Radiation exposure with use of intraoperative fluoroscopy is a potential orthopedic occupational risk factor. The purpose of this study was to perform a prospective comparison of hand versus eye radiation exposure associated with mini-C-arm utilization and to test the hypothesis that routine mini-C-arm does not yield hand or eye dosages exceeding current recommended levels. Over a 12-month period, hand and eye radiation exposure was prospectively measured in a single board-certified hand surgeon using mini-C-arm fluoroscopy. Twenty-five cases were performed utilizing mini-C-arm fluoroscopy. Average monthly hand radiation dosage (45.81±14.49 mrem) was significantly higher (p = .01) than eye radiation dosage (<30 mrem). Both recorded values were below their respective critical exposure limits, as reported by the International Commission on Radiological Protection. The findings suggest that hand and eye exposure associated with mini-C-arm utilization during routine surgery does not approach reported levels of critical radiation loads.
منابع مشابه
Hand and body radiation exposure with the use of mini C-arm fluoroscopy.
PURPOSE To determine whole body and hand radiation exposure to the hand surgeon wearing a lead apron during routine intraoperative use of the mini C-arm fluoroscope. METHODS Four surgeons (3 hand attending surgeons and 1 hand fellow) monitored their radiation exposure for a total of 200 consecutive cases (50 cases per surgeon) requiring mini C-arm fluoroscopy. Each surgeon measured radiation ...
متن کاملRadiation exposure with use of the mini-C-arm for routine orthopaedic imaging procedures.
BACKGROUND The use of mobile fluoroscopic devices during orthopaedic procedures is associated with substantial concern with regard to the radiation exposure to surgeons and support staff. The perceived increased risks associated with large c-arm devices have been well documented. However, no study to date has documented the relative radiation risk associated with the use of a mini-c-arm device....
متن کاملPatient and surgeon radiation exposure: comparison of standard and mini-C-arm fluoroscopy.
BACKGROUND Use of c-arm fluoroscopy is common in the operating room, outpatient clinic, and emergency department. Consequently, there is a concern regarding radiation exposure. Mini-c-arm fluoroscopes have gained popularity; however, few studies have quantified exposure during mini-c-arm imaging of a body part larger than a hand or wrist. The purpose of this study was to measure radiation expos...
متن کاملExposure to direct and scatter radiation with use of mini-C-arm fluoroscopy.
BACKGROUND Mini-c-arm fluoroscopy has become an important resource to the orthopaedic surgeon. Exposure of the orthopaedic surgical team to radiation during standard large-c-arm fluoroscopy has been well studied; however, little is known about the amount of exposure to which a surgical team is subjected with the use of mini-c-arm fluoroscopy. Moreover, there is controversy regarding the use of ...
متن کاملComparison of Ocular Radiation Exposure Utilizing Three Types of Leaded Glasses.
STUDY DESIGN Anthropomorphic phantoms were used to measure radiation exposure to the surgeon phantom's eye. Groups analyzed were: Group 1-no glasses (None); Group 2-leaded lenses without lead sides (WOLS); Group 3-leaded lenses with lead sides (WLS); and Group 4-sport wraparound leaded glasses (Sport). Glasses were 0.75 mm lead equivalent. OBJECTIVE To evaluate the efficacy of three types of ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of surgical orthopaedic advances
دوره 26 2 شماره
صفحات -
تاریخ انتشار 2018