External fixator configurations in tibia fractures: 1D optimization and 3D analysis comparison

نویسندگان

  • Luís M. Roseiro
  • M. Augusta Neto
  • Ana Amaro
  • Rogério P. Leal
  • Miguel C. Samarra
چکیده

The use of external fixation devices in orthopedic surgery is very common in open tibial fractures. A properly applied fixator may improve the healing process while one improperly applied might delay the healing process. The several external fixator systems used in clinical today, can be categorized into uniplanar-unilateral, uniplanar-bilateral, biplanar and multiplanar. The stability on the fracture focus and, therefore, the fracture healing process, is related with the type of external fixator configuration that is selected. The aim of this study is to discuss the principles for the successful application of unilateral-uniplanar external fixation, the assembly of its components, for the case of a transverse fractures using computational models. In this context, the fixation stiffness characteristics are evaluated using a simplified 1D finite element model for the tibia and external fixator. The beams are modeled with realistic cross-sectional geometry and material properties instead of a simplified model. The VABS (the Variational Asymptotic Beam Section analysis) methodology is used to compute the cross-sectional model for the generalized Timoshenko model, which was embedded in the finite element solver FEAP. The use of Timoshenko beam theory allows accounting for several kinds of loads, including torsion moments. Optimal design is performed with respect to the assembly of fixator components using a genetic algorithm. The optimization procedure is based on the evaluation of an objective function, which is dependent on the displacement at the fracture focus. The initial and optimal results are compared by performing a 3D analysis, for which different three-dimensional finite element models are created. The geometrical model of a tibia is created on the basis of data acquired by CAT scan, made for a healthy tibia of a 22 year old male. The 3D comparison of the 1D optimal results show a clear improvement on the objective function for the several load cases and, therefore, it is shown that appropriate selection of the external fixator geometrical features can lead to an improvement on the stability of the external fixator. The results obtained show that the optimal position of the side beam and the first pin should be as close as possible to the bone interface and as close as possible to the fracture focus, respectively. Concerning the second pin, it should be placed away from the first pin in case of flexion loads, to axial and torsion loads the second pin should be placed near the first pin.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Simulation and Optimization of Fracture Healing

The objective of this paper is to present an optimization algorithm for the optimization of the axial stiffness of a fixator used in fracture healing. To this end, numerical simulations of fracture healing with 2D and 3D models are used. The simulation is extended by a tool to observe the bridging of two bone fragments, called bridge detector. Using this bridge detector, the optimization criter...

متن کامل

Treatment of Pediatric Open Femoral Fractures with External Fixator Versus Flexible Intramedullary Nails

  Background: In children, inappropriate treatment of open femoral fractures may induce several complications. A few studies have compared the external fixator with flexible intramedullary nails in high-grade open femoral fractures of children. The present study aims at comparing results of these two treatment methods in open femoral fractures. Methods: In this descriptive analytical study, 27 ...

متن کامل

Ilizarov Versus AO External Fixator for the Treatment of Tibia Open Fractures

BACKGROUND In developing countries, Ilizarov or AO external fixator is usually used for treatment of tibial open fractures. The purpose of this study was to compare the efficacy of these two methods for treatment of tibial open fractures. METHODS From April 2002 to April 2010, 120 patients with open tibial fractures admitted to the Department of Orthopedics of Babol University of Medical Scie...

متن کامل

Unstable proximal extraarticular tibia fractures: a biomechanical evaluation of four methods of fixation.

OBJECTIVE To compare the biomechanical stability of extraarticular proximal tibia fractures reconstructed using a double-plate construct, locking plate system, hybrid external fixator, and single lateral periarticular plate, all from the same manufacturer. DESIGN Standardized proximal tibial fractures (AO classification 41-A3.2 and A3.3) in synthetic tibiae were stabilized using one of the fo...

متن کامل

Management of Tibial Metaphyseal Fractures by Hybrid External Fixator

Introduction: The Hybrid External Fixator combines the advantages of the monolateral pin fixators and the circular Ilizarov wire fixators. Principle of early motion has been touted as the functional savior of major intra articular injuries in application of hybrid external fixation. Materials & Methodology: 48 patients were considered for the study with tibial fractures admitted in Mamata Gener...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Computer methods and programs in biomedicine

دوره 113 1  شماره 

صفحات  -