Leakage evaluation of original and compatible implant–abutment connections: In vitro study using Rhodamine B
نویسندگان
چکیده
Leakage has been addressed as a major contributing factor to inflammatory reactions at the implant-abutment connection, leading to problems such as oral malodor, inflammation, and marginal bone loss. The aim of this study was to investigate in vitro the leakage at implant-abutment interface of OsseoSpeed™ implants connected to original and compatible abutments. A total of 28 OsseoSpeed implants were divided into four groups (n = 7). Each group was connected to four different abutments according to manufacturers' recommendations: group A (TiDesign™); group B (Natea™); group C (Dual™); and group D (Implanet™) abutments. The inner volume of each implant-abutment combination was calculated and leakage was detected for each group with spectrophotometric analysis at 1 h (D0) and 48 h (D1) of incubation time using Rhodamine B. At 1 h, leakage volume was significantly lower in TiDesign and Dual than in Natea and Implanet (P < 0.001). At 48 h, however, leakage was significantly lower between TiDesign and all other systems (P < 0.005). Compatible abutments do not fit internal connection of OsseoSpeed implants perfectly, which increases the leakage of the final assembly.
منابع مشابه
In Vitro Evaluation of Leakage at Implant-Abutment Connection of Three Implant Systems Having the Same Prosthetic Interface Using Rhodamine B
Objectives. Hollow space between implant and abutment may act as reservoir for commensal and/or pathogenic bacteria representing a potential source of tissue inflammation. Microbial colonization of the interfacial gap may ultimately lead to infection and bone resorption. Using Rhodamine B, a sensitive fluorescent tracer dye, we aim in this study to investigate leakage at implant-abutment connec...
متن کاملEvaluation of the Effect of Zirconia and Titanium Abutments on Microleakage of Implant-Abutment Interface Under Oblique Cyclic Loading In Vitro
Background and Aim: Oral bacteria can proliferate in the implant-abutment interface (IAI) and cause inflammation in the peri-implant tissues and adjacent bone. This study aimed to assess the effect of zirconia and titanium abutments on the microleakage of the IAI under oblique cyclic loading conditions. Materials and Methods: In this in-vitro study, 12 implant-abutment assemblies with zirconia...
متن کاملEffect of GapSeal® as a Sealing Material on Microgap and Microleakage at External Hexagon Implant Connections Following Cyclic Loading: An In Vitro Study
Background and Aim: The mismatch of the implant-abutment connection can produce instant stress and microleakage which result in mechanical and biological complications. This study aimed to investigate the influence of GapSeal® as a sealing material on the extent of microgap and microleakage at the external hexagon implant platform following cyclic loading. Materials and Methods: Sixteen implan...
متن کاملComparative Evaluation of Microleakage at the Interface of Titanium and Zirconium Abutments Following Oblique Cyclic Loading: An in vitro Study
Abstract Background and aim: Oral microbiota could proliferate the microspace between the implant and abutment, thereby cause inflammation in the peri-implant tissues and adjacent bone. This study aimed to investigate the effect of two types of abutments (zirconia and titanium) on microleakage at implant-abutment interface area under oblique cyclic loading in vitro. Materials and methods:In thi...
متن کاملComparative Evaluation of Microgap at the Interface of Titanium and Zirconium Abutments Following Oblique Cyclic Loading: An in vitro Study
Abstract Background and Aim : Microgap in the implant-abutment interface is one of the main challenges in the treatment of two-piece implants. This study aimed to investigate the effect of two types of abutments (zirconia and titanium) on microgap at implant-abutment interface area under oblique cyclic loading in vitro. Methods and Materials In this in vitro study, 12 implant-abutment assemb...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 5 شماره
صفحات -
تاریخ انتشار 2014