Once Annealed Highly Cross-Linked Polyethylene Exhibits Low Wear at 9 to 15 Years.
نویسندگان
چکیده
A once annealed highly cross-linked polyethylene (HXLPE) was introduced in 1998. Concerns regarding its long-term performance and oxidative resistance exist because of the presence of retained free radicals. The authors studied 48 patients with 50 hip implants having an average age of 62 years. They were followed for 9 to 15 years. The purpose of this study was to determine linear wear rate and the incidence of osteolysis and/or mechanical failure. At an average follow-up of 12.2 years, the annual linear wear rate was 0.018 mm (SD, 0.024 mm). No mechanical failures or osteolysis have been found to date. The clinical performance of this HXLPE continues to meet expectations despite the presence of free radicals. [Orthopedics. 2016; 39(3):e565-e571.].
منابع مشابه
Sequentially Annealed Highly Cross-linked Polyethylene Reduced In Vivo Wear Particle Generation In TKA
Introduction: Polyethylene wear particles induce macrophages to release cytokines, which can lead to osteolysis and aseptic loosening of total joint prostheses. The generation of polyethylene wear particles is one of the most important factors that affect the midterm and long-term clinical results associated with total knee arthroplasty (TKA). Therefore, to achieve better long-term results for ...
متن کاملIn Vivo Wear Particles of Annealed Highly Cross-Linked Polyethylene in Total Hip Arthropalsty
INTRODUCTION: Polyethylene wear particles induce cytokine release by macrophage, which leads to osteolysis and aseptic loosening in total joint arthroplasties. Generation of polyethylene wear particles is one of the most important factors that affecting mid-term and long-term results of total hip arthroplasty (THA). Therefore, to reduce wear generation and to achieve better long-term results of...
متن کاملNo adverse effects of submelt-annealed highly crosslinked polyethylene in cemented cups
BACKGROUND AND PURPOSE Highly crosslinked polyethylene (PE) is in standard use worldwide. Differences in the crosslinking procedure may affect the clinical performance. Experimenatal data from retrieved cups have shown free radicals and excessive wear of annealed highly crosslinked PE. We have previously reported low wear and good clinical performance after 6 years with this implant, and now re...
متن کاملCross-linked compared with historical polyethylene in THA: an 8-year clinical study.
UNLABELLED Wear particle-induced osteolysis is a major cause of aseptic loosening in THA. Increasing wear resistance of polyethylene (PE) occurs by increasing the cross-link density and early reports document low wear rates with such implants. To confirm longer-term reductions in wear we compared cross-linked polyethylene (irradiation in nitrogen, annealing) with historical polyethylene (irradi...
متن کاملHip Resurfacing Using Highly Cross-linked Polyethylene: Prospective Study Results at 8.5 Years.
BACKGROUND Hip resurfacing is an option to consider when treating younger, more active patients. Advantages over total hip arthroplasty include a more normal gait and a lower incidence of thigh pain. METHODS In this prospective study, 190 hip resurfacing procedures (164 participants) were performed using a cobalt-chromium femoral component and a cementless acetabular cup with a 3.8-mm highly ...