Gelsomino 1 Assessment of a Continuous Blood Gas Monitoring System in Animals during Circulatory stress. Short Title: Continuous Blood Gas Monitoring
نویسندگان
چکیده
Introduction: The study was aimed to determine the measurement accuracy of The CDI TM blood parameter monitoring system 500 (Terumo Cardiovascular Systems Corporation, Ann Arbor MI) in the real-time continuous measurement of arterial blood gases under different cardiocirculatory stress conditions Methods: Inotropic stimulation (Dobutamine 2.5 and 5 μg/kg/min), vasoconstriction (Argininevasopressin 4, 8 and 16 IU/h), hemorrhage (–10%, –20%, –35%, and –50% of the theoretical volemia), and volume resuscitation were induced in ten swine (57.4 ± 10.7Kg). Intermittent blood gas assessments were carried out using a routine gas analyzer at any experimental phase and compared with values obtained at the same time settings during continuous monitoring with CDI TM 500 system. The Bland-Altman analysis was employed. Results: Bias and precision for pO2 were − 0.06 kPa and 0.22 kPa, respectively (r 2 = 0.96); pCO2 − 0.02 kPa and 0.15 kPa, respectively; pH -0.001 and 0.01 units, respectively ( r 2 =0.96). The analysis showed very good agreement for SO2 (bias 0.04,precision 0.33, r 2 =0.95), Base excess (bias 0.04,precision 0.28, r 2 =0.98), HCO3 (bias 0.05,precision 0.62, r 2 =0.92),hemoglobin (bias 0.02,precision 0.23, r 2 =0.96) and K + (bias 0.02, precision 0.27, r 2 =0.93). The sensor was reliable throughout the experiment during hemodynamic variations. Conclusions: Continuous blood gas analysis with the CDI TM 500 system was reliable and it might represent a new useful tool to accurately and timely monitor gas exchange in critically ill patients. Nonetheless, our findings need to be confirmed by larger studies to prove its reliability in the clinical setting.
منابع مشابه
Assessment of a continuous blood gas monitoring system in animals during circulatory stress
BACKGROUND The study was aimed to determine the measurement accuracy of The CDI™ blood parameter monitoring system 500 (Terumo Cardiovascular Systems Corporation, Ann Arbor MI) in the real-time continuous measurement of arterial blood gases under different cardiocirculatory stress conditions METHODS Inotropic stimulation (Dobutamine 2.5 and 5 μg/kg/min), vasoconstriction (Arginine-vasopressin...
متن کاملContinuous fiberoptic arterial and venous blood gas monitoring in hemorrhagic shock.
STUDY OBJECTIVE To compare the performance of continuous fiberoptic blood gas monitoring with standard, intermittent blood gas sampling in the measurement of arterial and central venous blood gases during marked hemodynamic changes. DESIGN Prospective, consecutive, enrollment, experimental study. SETTING Research laboratory at a university medical center. PARTICIPANTS Seven anesthetized, ...
متن کاملA System for Continuous Estimating and Monitoring Cardiac Output via Arterial Waveform Analysis
Background: Cardiac output (CO) is the total volume of blood pumped by the heart per minute and is a function of heart rate and stroke volume. CO is one of the most important parameters for monitoring cardiac function, estimating global oxygen delivery and understanding the causes of high blood pressure. Hence, measuring CO has always been a matter of interest to researchers and clinicians. Sev...
متن کاملClinical assessment of a continuous intraarterial blood gas monitoring system.
We evaluated the clinical performance of a continuous intraarterial blood gas monitoring (CIABG) system which includes a fluorometric intravascular sensor. Seventeen patients undergoing elective surgery were monitored perioperatively with the CIABG system (PB3300; Puritan Bennett, Carlsbad, CA). Conventional laboratory blood gas analyses (BGA) were performed simultaneously whenever indicated, a...
متن کاملDesign and Evaluation of a Pressure and Temperature Monitoring System for Pressure Ulcer Prevention
Introduction Pressure ulcers are tissue damages resulting from blood flow restriction, which occurs when the tissue is exposed to high pressure for a long period of time. These painful sores are common in patients and elderly, who spend extended periods of time in bed or wheelchair. In this study, a continuous pressure and temperature monitoring system was developed for pressure ulcer preventio...