Frozen human serum reference material for standardization of sodium and potassium measurements in serum or plasma by ion-selective electrode analyzers.

نویسندگان

  • P C Gunaratna
  • W F Koch
  • R C Paule
  • A D Cormier
  • P D'Orazio
  • N Greenberg
  • K M O'Connell
  • A Malenfant
  • A O Okorodudu
  • R Miller
چکیده

Three interlaboratory round-robin studies (RR1, RR2, and RR3) were conducted to identify a serum-based reference material that would aid in the standardization of direct ion-selective electrode (ISE) measurements of sodium and potassium. Ultrafiltered frozen serum reference materials requiring no reconstitution reduced between-laboratory variability (the largest source of imprecision) more than did other reference materials. ISE values for RR3 were normalized by the use of two points at the extremes of the clinical range for sodium (i.e., 120 and 160 mmol/L), with values assigned by the flame atomic emission spectrometry (FAES) Reference Method. This FAES normalization of ISE raw values remarkably improved all sources of variability and unified the results from seven different direct ISE analyzers to the FAES Reference Method value. Subsequently, a three-tiered, fresh-frozen human serum reference material was produced to the specifications developed in RR1-RR3, was assigned certified values for sodium and potassium by Definitive Methods at the National Institute of Standards and Technology (NIST), and was made available in 1990 to the clinical laboratory community as a Standard Reference Material (SRM); it is now identified as SRM 956. Albeit retrospectively, we show how applying an FAES normalization step identical to that used in RR4/5 to the ISE data for SRM 956 after the NIST Definitive Method values were known, consistently moved the ISE results for RR3 closer to the true value for Na+ and K+.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Discrepancies between sodium concentrations measured by the Kodak Ektachem 700 and by dilutional and direct ion-selective electrode analyzers.

We have identified rare (approximately 0.2% of all samples), but clinically significant, discrepancies between serum or plasma sodium concentrations measured with the Kodak Ektachem 700's direct ion-selective electrode (ISE) method and concentrations measured with two other analyzers: the Beckman Synchron CX3's dilutional ISE instrument and the Radiometer KNA2 instrument for sodium-potassium an...

متن کامل

Determination of Copper Content of Human Blood Plasma by an Ion Selective Electrode based on a New Copper-Selectophore

A new selectophore was introduced for Cu2+ ions. Spectroscopic studies showed a selectivity of a new organic compound (L) toward copper ions and several transitional metal ions. Hence, L was utilized in designing several ion selective electrodes for these cations. In practice, Cu2+ ion selective electrode behaved Nernstian (slope of 27.95±0.3 mV decade-1) over a...

متن کامل

Chromogenic ionophore-based methods for spectrophotometric assay of sodium and potassium in serum and plasma.

We report spectrophotometric methods for quantifying sodium and potassium in serum and plasma without sample pretreatment or solvent-extraction steps. The methods are based on novel chromogenic ionophores characterized by molecular structures highly preorganized for binding, which result in exceptionally high sensitivity and selectivity (negligible responses to sodium at 80 to 200 mmol/L in the...

متن کامل

Benzalkonium interference with test methods for potassium and sodium.

Six automated instruments that measure sodium and potassium were tested for interference from two compounds used in catheters. Tridodecylmethylammonium heparin did not interfere with any of the methods. However, benzalkonium heparin falsely increased sodium measurement with the Kodak Ektachem, and falsely increased potassium measurements with three instruments (Beckman Astra, Baxter Paramax, an...

متن کامل

Measurement of ionized calcium in serum with ion-selective electrodes: a mature technology that can meet the daily service needs.

This article reviews key advances in ion-selective electrode technology that have made potentiometric measurements of ionized calcium (Ca2+) reliable and precise. Our use of two second-generation Ca2+ analyzers (Radiometer ICA1 and NOVA 8) made possible uninterrupted service as volume increased to 31 640 patient tests in 1985. The lower results on the NOVA 8 were adjusted upwards to match those...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Clinical chemistry

دوره 38 8 Pt 1  شماره 

صفحات  -

تاریخ انتشار 1992