A closed-loop artificial pancreas based on risk management.

نویسندگان

  • Fraser Cameron
  • B Wayne Bequette
  • Darrell M Wilson
  • Bruce A Buckingham
  • Hyunjin Lee
  • Günter Niemeyer
چکیده

BACKGROUND Control algorithms that regulate blood glucose (BG) levels in individuals with type 1 diabetes mellitus face several fundamental challenges. Two of these are the asymmetric risk of clinical complications associated with low and high glucose levels and the irreversibility of insulin action when using only insulin. Both of these nonlinearities force a controller to be more conservative when uncertainties are high. We developed a novel extended model predictive controller (EMPC) that explicitly addresses these two challenges. METHOD Our extensions to model predictive control (MPC) operate in three ways. First, they explicitly minimize the combined risk of hypoglycemia and hyperglycemia. Second, they integrate the effect of prediction uncertainties into the risk. Third, they understand that future control actions will vary if measurements fall above or below predictions. Using the University of Virginia/Padova Simulator, we compared our novel controller (EMPC) against optimized versions of a proportional-integral-derivative (PID) controller, a traditional MPC, and a basal/bolus (BB) controller, as well as against published results of an independent MPC (IMPC). The BB controller was optimized retrospectively to serve as a bound on the possible performance. RESULTS We tuned each controller, where possible, to minimize a published blood glucose risk index (BGRI). The simulated controllers (PID/MPC/EMPC/BB) provided BGRI values of 2.99/3.05/2.51/1.27 as compared to the published IMPC BGRI value of 4.10. These correspond to 73/79/84/92% of BG values lying in the euglycemic range (70-180 mg/dl), respectively, with mean BG levels of 151/156/147/140 mg/dl. CONCLUSION The EMPC strategy extends MPC to explicitly address the issues of asymmetric glycemic risk and irreversible insulin action using estimated prediction uncertainties and an explicit risk function. This controller reduces the avoidable BGRI by 56% (p < .05) relative to a published MPC algorithm studied on a similar population.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Closed-loop for type 1 diabetes – an introduction and appraisal for the generalist

BACKGROUND Rapid progress over the past decade has been made with the development of the 'Artificial Pancreas', also known as the closed-loop system, which emulates the feedback glucose-responsive functionality of the pancreatic beta cell. The recent FDA approval of the first hybrid closed-loop system makes the Artificial Pancreas a realistic therapeutic option for people with type 1 diabetes. ...

متن کامل

Algorithms for a Single Hormone Closed-Loop Artificial Pancreas: Challenges Pertinent to Chemical Process Operations and Control

The development of a closed-loop artificial pancreas to regulate the blood glucose concentration of individuals with type 1 diabetes has been a focused area of research for over 50 years, with rapid progress during the past decade. The daily control challenges faced by someone with type 1 diabetes include asymmetric objectives and risks, and one-sided manipulated input action with frequent rela...

متن کامل

Hypoglycemia Prevention in Closed-Loop Artificial Pancreas for Patients with Type 1 Diabetes

The current chapter addresses the problem of hypoglycemia in type 1 diabetes from biomedical and control engineering points of view. It gives a general introduction to the artificial pancreas system, and the risk of hypoglycemia in closed-loop insulin treatment. Then, it provides a review on the state of the art in hypoglycemia control, and the recent approaches in dealing with hypoglycemia in ...

متن کامل

Control of Flexible Link Robot using a Closed Loop Input-Shaping Approach

This paper is has addressed the Single Flexible Link Robot. The dynamical model is derived using Euler-Lagrange equation and then a proper controller is designed to suppress a  vibration based-on Input-Shaping (IS) method. But, IS control method is an open loop strategy. Due to the weakness of open loop control systems, a closed loop IS control system is proposed. The achieved closed loop c...

متن کامل

Continuous glucose monitoring considerations for the development of a closed-loop artificial pancreas system.

BACKGROUND Commercialization of a closed-loop artificial pancreas system that employs continuous subcutaneous insulin infusion and interstitial fluid glucose sensing has been encumbered by state-of-the-art technology. Continuous glucose monitoring (CGM) devices with improved accuracy could significantly advance development efforts. However, the current accuracy of CGM devices might be adequate ...

متن کامل

Development of closed-loop supply chain mathematical model (cost-benefit-environmental effects) under uncertainty conditions by approach of genetic algorithm

In the current world, the debate on the reinstatement and reuse of consumer prod-ucts has become particularly important. Since the supply chain of the closed loop is not only a forward flow but also a reverse one; therefore, companies creating integ-rity between direct and reverse supply chain are successful. The purpose of this study is to develop a new mathematical model for closed loop suppl...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of diabetes science and technology

دوره 5 2  شماره 

صفحات  -

تاریخ انتشار 2011