Late-onset right ventricular failure in patients with preoperative small left ventricle after implantation of continuous flow left ventricular assist device.

نویسندگان

  • Teruhiko Imamura
  • Koichiro Kinugawa
  • Naoko Kato
  • Hironori Muraoka
  • Takeo Fujino
  • Toshiro Inaba
  • Hisataka Maki
  • Osamu Kinoshita
  • Masaru Hatano
  • Shunei Kyo
  • Minoru Ono
چکیده

BACKGROUND  The continuous flow (CF) left ventricular assist device (LVAD) has replaced the pulsatile flow (PF) LVAD because of its advantages of better patient survival and higher quality of life. However, "late-onset right ventricular failure (RVF)" after CF LVAD implantation has emerged as an increasing concern, but little is known about the mechanism. METHODS AND RESULTS  We retrospectively analyzed the 3-month hemodynamic and echocardiographic data from 38 consecutive patients who had received CF LVADs, and from 22 patients who had received PF LVADs. Late-onset RVF was defined as persistent right ventricular stroke work index (RVSWI) <4.0g/m(2) at any rotation speed and after saline infusion test at 5 weeks after implantation of CF LVAD. Patients with late-onset RVF had significantly impaired exercise tolerance indicated by shorter 6-min walking distance and lower peak VO2, and worsened tricuspid regurgitation, together with enlargement of the RV under CF LVAD treatment (all P<0.05). Univariable analyses demonstrated that preoperative smaller LV diastolic diameter (LVDd) was the risk factor for late-onset RVF with a cutoff value of 64mm calculated by ROC analysis (area under curve, 0.925). In contrast, there was no correlation between preoperative LVDd and postoperative RVSWI in the PF LVAD group, though their preoperative background was worse than that of the CF group. CONCLUSIONS  In the setting of preoperative small LVDd, CF LVAD may cause late-onset RVF by leftward shift of the interventricular septum.  (Circ J 2014; 78: 625-633).

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Right ventricular failure in patients with the HeartMate II continuous-flow left ventricular assist device: incidence, risk factors, and effect on outcomes.

OBJECTIVE The aim of this study was to evaluate the incidence, risk factors, and effect on outcomes of right ventricular failure in a large population of patients implanted with continuous-flow left ventricular assist devices. METHODS Patients (n = 484) enrolled in the HeartMate II left ventricular assist device (Thoratec, Pleasanton, Calif) bridge-to-transplantation clinical trial were exami...

متن کامل

Improvement of Left Ventricular Assist Device (LVAD) in Artificial Heart Using Particle Swarm Optimization

In this approach, the Left ventricular assist pump for patients with left ventricular failure isused. The failure of the left ventricle is the most common heart disease during these days. Inthis article, a State feedback controller method is used to optimize the efficiency of a samplingpump current. Particle Swarm Algorithm, which is a set of rules to update the position andvelocity, is applied...

متن کامل

Right ventricular dysfunction following continuous flow left ventriccular assist device placement in 51 patients: predicators and outcomes

BACKGROUND Right ventricular (RV) dysfunction following implantation of a left ventricular assist device (LVAD) is a serious condition and is associated with increased mortality. METHODS The aim of the study is to investigate the significance of pre-existing RV dysfunction, tricuspid valve (TV) insufficiency, and the severity of septal deviation following LVAD implantation on RV dysfunction, ...

متن کامل

Application of centrifugal pump in right ventricular failure due to pulmonary artery hypertension. (A case report)

Background: It has not been yet developed a ost suitable method to treat right ventricular failure due to pulmonar artery hypertention resulted from ventricular arrhythia. Though some case reports about patients ready to heart transplant using left assist device or biventral assist device as bridge have been published in Journals, but nocase concerning application of centrifugal pump for treatm...

متن کامل

Left ventricular assist devices decrease fixed pulmonary hypertension in cardiac transplant candidates.

OBJECTIVE Fixed pulmonary hypertension is a contraindication for cardiac transplantation because of the increased risk of donor heart failure. We sought to determine whether left ventricular assist devices improve fixed pulmonary hypertension in cardiac transplant candidates to enable safe cardiac transplantation. METHODS Thirty-five consecutive cardiac transplant candidates (age 56 +/- 6 yea...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Circulation journal : official journal of the Japanese Circulation Society

دوره 78 3  شماره 

صفحات  -

تاریخ انتشار 2014