BACKGROUND AND OBJECTIVES The number of patients with cardiac implantable electronic devices needing lead extraction is increasing for various reasons, including infections, vascular obstruction, and lead failure. We report our experience with transvenous extraction of pacemaker and defibrillator leads via the inferior approach of using a gooseneck snare as a first-line therapy and compare extraction using a gooseneck snare with extraction using simple manual traction. SUBJECTS AND METHODS The study included 23 consecutive patients (43 leads) who underwent transvenous lead extraction using a gooseneck snare (group A) and 10 consecutive patients (17 leads) who underwent lead extraction using simple manual traction (group B). Patient characteristics, indications, and outcomes were analyzed and compared between the groups. RESULTS The dwelling time of the leads was longer in group A (median, 121) than in group B (median, 56; p=0.000). No differences were noted in the overall procedural success rate (69.6% vs. 70%), clinical procedural success rate (82.6% vs. 90%), and lead clinical success rate (86% vs. 94.1%) between the groups. The procedural success rates according to lead type were 89.2% and 100% for pacing leads and 66.7% and 83.3% for defibrillator leads in groups A and B, respectively. Major complications were noted in 3 (mortality in 1) patients in group A and 2 patients in group B. CONCLUSION Transvenous extraction of pacemaker leads via an inferior approach using a gooseneck snare was both safe and effective. However, stand-alone transvenous extraction of defibrillator leads using the inferior approach was suboptimal.
منابع مشابه
Lead Extraction: "Drag Through" Technique
The goal of extraction techniques of chronic pacemaker and defibrillator leads is to present an approach that is successful in extracting all leads and minimizes or eliminates complications. To extract a chronically implanted intravascular device, the device must be separated from the encapsulating inflammatory tissue. Three ablation techniques are currently used: mechanical, laser and electros...
متن کاملSuccessful extraction of an implantable cardioverter-defibrillator lead in a patient with pocket infection via the femoral approach with a basket snare
A 27-year-old man was admitted to our institution with implantable cardioverter-defibrillator (ICD) pocket infection. The ICD (single chamber, dual shock coils) was implanted for the prevention of aborted sudden cardiac death due to idiopathic ventricular fibrillation. Because we could not use locking stylet and countertraction manuever, baseket snare with metal wire should be hanged at the max...
متن کاملRemoval of the broken part of implantable cardioverter-defibrillator's electrode causing pulmonary embolism via femoral vein.
A 66-year-old patient with idiopathic dilated cardiomyopathy underwent transvenous extraction of an implantable cardioverter-defibrillator. The distal part of the electrode was broken during manual traction through the left subclavian vein. In the present case, we showed a rare complication of transvenous lead extraction and its management.
متن کاملImplantable defibrillator lead extraction with optimized standard extraction techniques
BACKGROUND Implantable cardioverter-defibrillator (ICD) leads might not be extracted especially in developing countries because of the high cost and lack of specialized tools. We aimed to evaluate transvenous extraction of ICD leads using optimized standard techniques. METHODS We prospectively analyzed clinical characteristics, optimized extraction techniques and the feasibility of extraction...
متن کاملUnexpected challenging case of coronary sinus lead extraction
An 84-year-old woman implanted with cardiac resynchronization therapy defibrillator underwent transvenous lead extraction 4 mo after the implant due to pocket infection. Atrial and right ventricular leads were easily extracted, while the attempt to remove the coronary sinus (CS) lead was unsuccessful. A few weeks later a new extraction procedure was performed in our center. A stepwise approach ...