Assessment of a 2D EPID-based Dosimetry Algorithm for Pre-treatment and In-vivo Midplane Dose Verification

نویسندگان

  • Ali Jomehzadeh Department of Medical Physics, Kerman University of Medical Sciences, Kerman, Iran. Department of Radiation Oncology, The Netherlands Cancer Institute, Amsterdam, The Netherlands.
  • Alireza Amouheidari Department of Radiation Oncology, Isfahan Milad Hospital, Isfahan, Iran.
  • Anton Mans Department of Radiation Oncology, The Netherlands Cancer Institute, Amsterdam, The Netherlands.
  • Ben Mijnheer Department of Radiation Oncology, The Netherlands Cancer Institute, Amsterdam, The Netherlands.
  • Mohammad Mohammadi .Medical Physics Department, Royal Adelaide Hospital. Adelaide, SA 5000, Australia
  • Parvaneh Shokrani Department of Medical Physics, Isfahan University of Medical Sciences, Isfahan, Iran.
چکیده مقاله:

Introduction: The use of electronic portal imaging devices (EPIDs) is a method for the dosimetric verification of radiotherapy plans both pretreatment and in-vivo. The aim of this study was to test a 2D EPID-based dosimetry algorithm for dose verification of some plans inside a homogenous and anthropomorphic phantom and in-vivo, as well. Materials and Methods: Dose distributions were reconstructed from EPID images using a 2D EPID dosimetry algorithm inside a homogenous slab phantom for a simple 10×10 cm2 box technique, the 3D conformal (prostate, head-and-neck and lung) and an IMRT prostate plans inside an anthropomorphic (Alderson) phantom and the patients (one fraction in-vivo) for the 3D conformal plans (prostate, head-and-neck and lung). Results: The Planned and EPID dose difference at isocenter, on average, was 1.7% for the pretreatment verification and was less than 3% for all in vivo plans except a head-and-neck which was 3.6%. The mean γ values for a 7-field prostate IMRT plan delivered to the Alderson phantom varied from 0.28 to 0.65. For 3D conformal plans applied for the Alderson phantom, all γ1% values were within the tolerance level for all plans and in both AP-PA beams. Conclusion: The 2D EPID-based dosimetry algorithm provides an accurate method to verify the dose of a simple 10×10 cm2 fields in two dimensions inside a homogenous slab phantom as well as for IMRT prostate plan and 3D conformal plans (prostate, head-and-neck and lung plans) applied using an anthropomorphic phantom and in-vivo.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification.

The aim of this study was to demonstrate how dosimetry with an amorphous silicon electronic portal imaging device (a-Si EPID) replaced film and ionization chamber measurements for routine pre-treatment dosimetry in our clinic. Furthermore, we described how EPID dosimetry was used to solve a clinical problem. IMRT prostate plans were delivered to a homogeneous slab phantom. EPID transit images w...

متن کامل

EPID in vivo Dosimetry

Introduction: The most modern radiotherapy devices are equipped with an Electronic Portal Imaging Device (EPID) system which is located on opposite side of the machine’s head. EPID system is often used to setting up the position verification during or between radiotherapy sessions. Material and Methods: Various types of dosimeters have been used to setting up ...

متن کامل

3D EPID based dosimetry for pre-treatment verification of VMAT – methods and challenges

This article presents an overview of pre-treatment verification of volumetric modulated arc therapy (VMAT) with electronic portal imaging devices (EPIDs). Challenges to VMAT verification with EPIDs are discussed including EPID sag/flex during rotation, acquisition using cine-mode imaging, image artefacts during VMAT and determining the gantry angle for each image. The major methods that have be...

متن کامل

3D EPID-based in vivo dosimetry for IMRT and VMAT

In this paper the various approaches of EPID-based in vivo IMRT and VMAT dose verification, and their clinical implementation, are described. It will be shown that EPID-based in vivo dosimetry plays an important role in the total chain of verification procedures in a radiotherapy department. EPID-based dosimetry, in combination with in-room imaging, is a fast and accurate tool for 3D in vivo ve...

متن کامل

Evaluation of a fast method of EPID-based dosimetry for IMRT and Comparison with 3D EPID-based dosimetry system using conventional two- and three-dimensional detectors for VMAT

Introduction: Electronic portal imaging devices (EPIDs) could potentially be useful for intensity-modulated radiation therapy (IMRT) and VMAT QA. The data density, high resolution, large active area, and efficiency of the MV EPID make it an attractive option. However, EPIDs were designed as imaging devices, not dosimeters, and as a result they do not inherently measure dose in ...

متن کامل

EPID‐based dosimetry to verify IMRT planar dose distribution for the aS1200 EPID and FFF beams

We proposed to perform a basic dosimetry commissioning on a new imager sys-tem, the Varian aS1200 electronic portal imaging device (EPID) and TrueBeam 2.0 linear accelerator for flattened (FF) and flattening filter-free (FFF) beams, then to develop an image-based quality assurance (QA) model for verification of the system delivery accuracy for intensity-modulated radiation therapy (IMRT) treat-...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ذخیره در منابع من قبلا به منابع من ذحیره شده

{@ msg_add @}


عنوان ژورنال

دوره 15  شماره Special Issue-12th. Iranian Congress of Medical Physics

صفحات  365- 365

تاریخ انتشار 2018-12-01

با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023