Comparison of Central Corneal Thickness Measured through Pentacam, OCTCasia2, and Orbscan IIz in Normal Individuals

نویسندگان

  • , Mashhad University of Medical Sciences, Mashhad, Iran
  • Azimi, A. Mashhad University of Medical Sciences, Mashhad, Iran
  • Bahrami, M. Mashhad University of Medical Sciences, Mashhad, Iran
  • Momeni, H. Mashhad University of Medical Sciences, Mashhad, Iran
  • Sedaghat , MR. Mashhad University of Medical Sciences, Mashhad, Iran
  • Tabesh, H. Mashhad University of Medical Sciences, Mashhad, Iran
چکیده مقاله:

Purpose: To compare the central corneal thickness measured through Pentacam, OCTCasia2, and Orbscan IIz in normal individuals. Methods: In this cross sectional comparative study, the researcher investigated the eyes of 70 normal candidates (140 eyes) with the mean age of 31.42 ± 7.70, referring to Dr. Sedaghat eye clinic for refractive surgery. For each person, Central Corneal Thickness (CCT) was measured, using 3 devices including Pentacam, OCTCasia2 and Orbscan. The data were collected and analyzed. The Pearson Correlation Coefficient was employed to assess correlation of 3 devices with each other. Bland-Altman plot was also used to evaluate the agreement between devices. Results: The mean CCT measured through Orbscan, Pentacam, and OCTCasia2 was reported to be 535.06±35.11 µm, 525.21±31.60 µm, and 511.06±30.87 µm, respectively. The Pearson Correlation Coefficient between Pentacam-OCTCasia2, OCTCasia2-Orbscan, Pentacam-Orbscan were r=0.955; P<0.001, r=0.913; P<0.001, r=0.941; P<0.001, respectively. Conclusion: There was a very good correlation and agreement between devices, but none of the devices had the ability to replace each other.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of Central Corneal Thickness Measurements with Pentacam, Orbscan II, and Ultrasound Pachymeter

Purpose: To compare thickness of central cornea measured using Pentacam, Orbscan II, and ultrasound pachymeter Methods: Patients with no history of corneal diseases or systemic diseases affecting eyes, who did not wear contact lens or use eye medications, and who with no previous history of corneal surgery were selected for this study. Central corneal thickness (CCT) was measured by three metho...

متن کامل

Central corneal thickness, anterior chamber depth, and pupil diameter measurements using Visante OCT, Orbscan, and Pentacam.

PURPOSE To evaluate the agreement among three different optical methods in measuring anterior chamber depth (ACD), central corneal thickness (CCT), and pupil diameter. METHODS One hundred eyes of 50 healthy patients (25 men, 25 women) were enrolled in the study. Mean patient age was 25 years (range: 21 to 32 years). Exclusion criteria were history of any intraocular or corneal surgery, contac...

متن کامل

Comparison of Central Corneal Thickness Measurements by Pentacam and Ultrasound Pachymetry in Normal Myopic Patients

INTRODUCTION Precise measurement of corneal thickness is of crucial importance in corneal refractive procedures. The aim of this work is to compare the central corneal thickness measurements using Scheimpflug imaging (Pentacam) and contact US pachymetry in normal myopic patients. METHODS This retrospective cohort study was conducted on in 144 eyes of 72 normal myopic patients during 6 months....

متن کامل

comparison of central corneal thickness measurements with pentacam, orbscan ii, and ultrasound pachymeter

purpose: to compare thickness of central cornea measured using pentacam, orbscan ii, and ultrasound pachymeter methods: patients with no history of corneal diseases or systemic diseases affecting eyes, who did not wear contact lens or use eye medications, and who with no previous history of corneal surgery were selected for this study. central corneal thickness (cct) was measured by three metho...

متن کامل

Agreement between Clinical History Method, Orbscan IIz, and Pentacam in Estimating Corneal Power after Myopic Excimer Laser Surgery

The purpose of this study was to investigate the agreement between the clinical history method (CHM), Orbscan IIz, and Pentacam in estimating corneal power after myopic excimer laser surgery. Fifty five patients who had myopic LASIK/PRK were recruited into this study. One eye of each patient was randomly selected by a computer-generated process. At 6 months after surgery, postoperative corneal ...

متن کامل

Measurement of Central Corneal Thickness Using Ultrasound Pachymetry and Orbscan II in Normal Eyes

PURPOSE To compare ultrasound pachymetry and Orbscan II for measurement of central corneal thickness (CCT) in normal eyes. METHODS The current study was performed at Labbafinejad Medical Center (LMC), Tehran, Iran. Three hundred eyes from 150 healthy individuals referred for keratorefractive surgery were assessed first by Orbscan II and then by ultrasound pachymetry, and CCT values were recor...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ذخیره در منابع من قبلا به منابع من ذحیره شده

{@ msg_add @}


عنوان ژورنال

دوره 24  شماره 2

صفحات  94- 98

تاریخ انتشار 2019-03

با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.

کلمات کلیدی

کلمات کلیدی برای این مقاله ارائه نشده است

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023