Dosimetric verification of pre-treatment intensity modulated radiation therapy in the commissioning process

نویسندگان

  • A. Ameri Radiation Oncology Department, Faulty of Medicine, Shahid Beheshti University of Medical Sciences, Tehran, Iran
  • A.R. Nikoofar Radiation Oncology Department, Faulty of Medicine, Iran University of Medical Sciences. Tehran, Iran
  • G. Esmaili Department of Medical Physics. Iran University of Medical Sciences, Tehran, Iran
  • P. Fadavi Radiation Oncology Department, Faulty of Medicine, Iran University of Medical Sciences. Tehran, Iran
  • S.R. Mahdavi Radiation Biology Research Center & Medical Physics Department, Faculty of Medicine, Iran University of Medical Sciences, Tehran, Iran
  • V. Nazari Department of Medical Physics, Faculty of Medicine, Hamedan University of Medical Sciences, Hamedan, Iran
چکیده مقاله:

Background: Commissioning tests are recommended before implementing Intensity Modulated Radiation Therapy. This study evaluated five different plans of IMRT mocks for different modulation indices. Materials and Methods: IMRT tests were done in slab phantom for both 6MV and 15 MV photon energies using a 2D array ion chamber dosimeter. Results: The acceptance criteria were based on 3% /3mm. All tests passed the gamma criteria except for the prostate test with 15 MV photon beams and C shape test for both 6 MV and 15 MV energies. The Ratio passing points for Multi target test in 6 MV and 15 MV plans were 93.1% and 91.5%, respectively. The results of the prostate tests with 3%/3mm was 92.3% for 6MV and 91.2% for 15 MV in 4%/4 mm tolerance limit. For Head and Neck test with the same gamma criteria, the percentages of the points were 93.2 and 94 for 6 MV and 15 MV plans, respectively. For C shape tests, the used gamma criteria were 4%/ 4 mm. The ratio passing points were 94.9% and 94.3% for 6MV and 15MV plans, respectively. However, C shape hard test could not pass the gamma criteria of 4%/ 4mm for 6MV. Conclusion: results showed that by increasing the complexity of the IMRT plan, the verification test must be done in a more strict-manner, because a small change in dose delivery can cause a large discrepancy between planned and real dosimetry and may produce hot spots in organs or a cold spot in the target volume.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Dosimetric verification of micro‐MLC based intensity modulated radiation therapy

A methodology for the dosimetric verification of micro-multileaf collimator MMLC) based intensity modulated radiation therapy (IMRT) plans intended for stereotactic applications is described. The procedure is similar to that of conventional IMRT patient-specific quality assurance with some notable exceptions. Relative dosimetry measurements are performed with radiographic film, a commercial fil...

متن کامل

Intensity modulated radiation therapy versus three‐dimensional conformal radiation therapy for the treatment of high grade glioma: a dosimetric comparison

The present study compared the dosimetry of intensity-modulated radiation therapy (IMRT) and three-dimensional conformal radiation therapy (3D-CRT) techniques in patients treated for high-grade glioma. A total of 20 patients underwent computed tomography treatment planning in conjunction with magnetic resonance imaging fusion. Prescription dose and normal-tissue constraints were identical for t...

متن کامل

Volumetric intensity-modulated arc therapy vs conventional intensity-modulated radiation therapy in nasopharyngeal carcinoma: a dosimetric study

Dosimetric comparisons between RapidArc (RA) and conventional Intensity-Modulated Radiation Therapy (IMRT) techniques for nasopharyngeal carcinoma (NPC) were performed to address differences in dose coverage of the target, sparing of organs-at-risk (OARs), delivery of monitor units (MUs) and time, to assess whether the RA technique was more beneficial for treatment of NPC. Eight NPC patients (S...

متن کامل

Intensity-Modulated Radiation Therapy

INSTRUCTIONS FOR USE This Medical Policy provides assistance in interpreting UnitedHealthcare benefit plans. When deciding coverage, the enrollee specific document must be referenced. The terms of an enrollee's document (e.g., Certificate of Coverage (COC) or Summary Plan Description (SPD)) may differ greatly. In the event of a conflict, the enrollee's specific benefit document supersedes this ...

متن کامل

Dosimetric evaluation of radiation dose rate effect in respiratory gated intensity modulated radiation therapy†

PURPOSE To investigate the dosimetric accuracy of the sliding window gated IMRT compared with the static treatment, using varying dose rates. MATERIALS AND METHODS This study measured changes in output and diode array response with changing dose rate, verified the precision of the motion table, and measured changes in dose distribution accuracy with film and diodes at two depths with changing...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ذخیره در منابع من قبلا به منابع من ذحیره شده

{@ msg_add @}


عنوان ژورنال

دوره 16  شماره 4

صفحات  493- 497

تاریخ انتشار 2018-10

با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.

کلمات کلیدی

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023