Estimate bowtie filter shape in PET/CT scan with TLD

نویسندگان

  • Alireza Karimian Associate Professor, Department of Biomedical Engineering, University of Isfahan, Isfahan, Iran
  • Arman Rahmim Associate Professor, Department of Radiology, Johns Hopkins University, Baltimore, Maryland, USA
  • Mahdi Nasri Nasrabadi Associate Professor, Department of Nuclear Engineering, Faculty of Advanced Sciences & Technologies, University of Isfahan, Isfahan, Iran, Tel.: + 98 31 3793 4222; Fax: + 98 31 3793 2342, [email protected]
  • Nematollah Ahmadi Jeshvaghane PhD candidate, Department of Nuclear Engineering, Faculty of Advanced Sciences & Technologies, University of Isfahan, Isfahan, Iran
چکیده مقاله:

Introduction: The CT machine utilizes a bowtie filter to shape the X-ray beam and remove lower energy photons. Configuration of this bowtie filter is complex and its geometry is often not available in detail. It causes the CT dose index (CTDI) be with the different values in measurement versus Monte Carlo simulation studies and other analytical calculations. It is important especially in dosimetry of internal organs. In this study, the bowtie filter shape is extracted by using Thermolumenecence dosimeters (TLDs).   Materials and Methods: In this work, the shape of the bowtie filter of the Biograph 6 PET/CT was derived by using 31 Thermolumenecence chips dosimeters (GR-200 series) made of LiF: Mg, Cu, and P were utilized. These chips calibrated by Cs-137 in SSDL organization and ECC (Element Correction Coefficient) of them by using TLD reader Harshaw-4000. Eventually, to evaluate the accuracy of the body bowtie filter shape as generated by TLDs, Monte Carlo simulation of CT was performed. 16 X-ray sources in various angles were used within the Monte Carlo code (MCNP-4C) to simulate the CT section of the PET-CT Biograph 6 system and to calculate dose. In the simulation code, absorbed dose was determined for CT scan with MeV/gram/source particle units and converted into absorbed dose with mGy/100 mAs units by conversion factor (CF).   Results: The relative difference (RD) between simulated and measured CTDI value for the PET/CT Siemens Biograph 6 at 80, 110 and 130kVp were 4.2, 2.9 and 2.3%, respectively.   Conclusion: In summarize The results showed that it is possible to determine the shape of the bowtie filter in PET/CT using an inexpensive TLDs with acceptable accuracy

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

stimate bowtie filter shape in PET/CT scan with TLD

Introduction: The CT machine utilizes a bowtie filter to shape the X-ray beam and remove lower energy photons. Configuration of this bowtie filter is complex and its geometry is often not available in detail. It causes the CT dose index (CTDI) be with the different values in measurement versus Monte Carlo simulation studies and other analytical calculations. It is important esp...

متن کامل

A TLD based method to estimate bowtie filter shape in PET/CT

Background: In CT systems, the machine utilizes a bowtie filter to shape the X-ray beam and remove lower energy photons. The shape of this bowtie filter is complicated and its geometry is often not available in detail. These renders the CT dose index (CTDI) to have different values in measurement versus Monte Carlo simulation studies and other analytical calculations especially in dosimetry of ...

متن کامل

Usability assessment of cone beam computed tomography with a full-fan mode bowtie filter compared to that with a half-fan mode bowtie filter

Background: In intensity modulated radiation therapy, cone beam computed tomography (CT) has been used to evaluate patients prior to treatment. This study conducted a comparative evaluation of the image reconstruction ability of the clinically used half-fan bowtie filter and the full-fan bowtie filter. Materals and Methods: A CT simulation marker was inserted inside a human phantom, and the pel...

متن کامل

Dynamic Bowtie Filter for Cone-Beam/Multi-Slice CT

A pre-patient attenuator ("bowtie filter" or "bowtie") is used to modulate an incoming x-ray beam as a function of the angle of the x-ray with respect to a patient to balance the photon flux on a detector array. While the current dynamic bowtie design is focused on fan-beam geometry, in this study we propose a methodology for dynamic bowtie design in multi-slice/cone-beam geometry. The proposed...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ذخیره در منابع من قبلا به منابع من ذحیره شده

{@ msg_add @}


عنوان ژورنال

دوره 15  شماره Special Issue-12th. Iranian Congress of Medical Physics

صفحات  395- 395

تاریخ انتشار 2018-12-01

با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023