Neutron Contamination Unexpected Dose in Varian LINAC

نویسندگان

  • Asghar Hadadi Assistant professor, Department of Nuclear Engineering, Science and Research Branch, Islamic Azad University, Tehran, Iran
  • Dariush Sardari Assistant professor, Department of Nuclear Engineering, Science and Research Branch, Islamic Azad University, Tehran, Iran
  • Elham Saeedzadeh Assistant professor, Department of Nuclear Engineering, Science and Research Branch, Islamic Azad University, Tehran, Iran
چکیده مقاله:

Introduction: Neutron contamination is very harmful for the patients receiving radiation therapy due to the high relative biological effect of neutrons. Medical linear accelerators head are made up of materials with high atomic numbers, and High-energy photons interaction with heavy materials can produce neutron contamination. In this study, neutron equivalent dose, neutron spectrum, and the effect of field size on neutron contamination dose is calculated by the FLUKA Monte Carlo code system. Materials and Methods: In this study, all parts of the Varian Clinac 2100 accelerator head’s elements, such as target, flattening filter, jaws, MLC, shields, primary and secondary collimators are simulated by the FLUKA code system. LINAC performance energy was considered 18MV and a water phantom with dimension of 140×140×35 cm3 was simulated for neutron dose calculation. To investigate the effect of field size on neutron contamination, larger treatment fields such as 10×10, 20×20, 30×30 and 40×40 was formed by the jaws, and the smaller ones like 1×1, 3×3 and 5×5 was formed by the MLC.   Results: The results show that the Neutron contamination dose can be up to 2.54 mSv per 1Gy photon dose at isocenter. Neutron to photon dose ratio is 0.004 at isocenter and more than 8.6 in out-of-field areas. Furthermore, the neutron dose changes at the surface of the water phantom is only 15% that show the protective elements of the LINAC head are not able to control the neutron contamination flux. The most of neutron dose distribution is at the surface of the water phantom, So it can increases the secondary cancer risk in patients, including skin cancer. The effect of treatment field size on neutron contamination shows that the larger field size formed by the jaws contribute to higher in-field neutron dose, and using of smaller field size formed by the MLC, cause the lower in-field neutron dose. Conclusion: The neutron contamination dose can not be neglected, and the effective parameters in neutron contamination exasperation should be thoroughly investigated. Furthermore, the harmful effects of the neutron contamination should be minimized by the appropriate protective shields

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Unexpected dose and the biological effect of neutron contamination in linear accelerators

Employing High-energy photons in medical linear accelerators can lead to neutron contamination. The relative biological effect of neutrons is 20 times higher than photons. Consequently, its damage to the biological tissue is much higher and unobtrusive. In this study, we calculated neutron equivalent dose in a medical linear accelerator by the FLUKA Monte Carlo code system. The results show tha...

متن کامل

Dental Materials Effect in Neutron Contamination: Electron Mode of a Linac

Background: Neutron contamination is produced in electron beams of linac when tooth or dental materials are located in the path of beam. Objective: This study aims to determine the neutron dose contamination from different dental restoration materials in electron mode of a linac. Material and Methods: In this experimental study, the neutron dose contamination was calculated in the p...

متن کامل

Fast Neutron Dose Mapping in a Linac Radiotherapy Facility

CR-39 plastic detectors were used for fast neutron dose mapping in the radiotherapy facility at King AbdulAziz University Hospital (KAUH). Detectors were calibrated using a Cf neutron source and a neutron dosimeter. After exposure chemical etching was performed using 6N NaOH solution at 70 C. Tracks were counted using an optical microscope and the number of tracks/cm was converted to a neutron ...

متن کامل

The influence of neutron contamination on pacemaker in photon beam radiotherapy by LINAC using the Monte Carlo method

  In radiation therapy with high-energy photon beams (E > 10 MV) neutrons are generated mainly in LINACs head thorough (γ,n) interactions. These neutrons affect the shielding requirements in radiation therapy rooms. According to the AAPMTG-34 report, photon absorbed dose of 10Gy can cause permanent damage to the pacemaker and the dose of 2Gy can make minor changes in the functioning of the pac...

متن کامل

The influence of neutron contamination on pacemaker in photon beam radiotherapy by LINAC using the Monte Carlo method

Introduction: In radiation therapy with high-energy photon beams (E > 7 MeV) neutrons are generated mainly in LINACs head thorough (γ, n) interactions. These neutrons affect the shielding requirements in radiation therapy rooms. According to the protocol TG-34, photon absorbed dose of 10Gy can cause permanent damage to the pacemaker and the dose of 2Gy can make minor changes in...

متن کامل

Using Zirconium Borohydride as Neutron Shield in LINAC head

Introduction: One of the concerns about using of high energy photons in radiotherapy is neutron contamination production. The threshold energy of photoneutron production in most materials with high atomic number is about 7MeV. Therefore, it seems to be logical that the main components with high atomic number elements used in the LINAC head have a large contribution to the neutr...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ذخیره در منابع من قبلا به منابع من ذحیره شده

{@ msg_add @}


عنوان ژورنال

دوره 15  شماره Special Issue-12th. Iranian Congress of Medical Physics

صفحات  138- 138

تاریخ انتشار 2018-12-01

با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023