Small photon field dosimetry using EBT2 Gafchromic film and Monte Carlo simulation

نویسندگان

  • H.A. Nedaie Department of Medical Physics , Faculty of Medicine, Tehran University of Medical Sciences, Tehran, Iran
  • Kh. Asnaashari Department of Radiotherapy , Cancer Institute, Tehran University of Medical Sciences, Tehran, Iran
  • M. Allahverdi Department of Medical Physics , Faculty of Medicine, Tehran University of Medical Sciences, Tehran, Iran
  • M. Yarahmadi Kurdistan University of Medical Sciences, Sanandaj, Iran
  • O.A. Sauer Klinik und Poliklinik für Strahlentherapie, University of Würzburg, Josef-Schneider-Strasse 11, 97080 Würzburg, Germany
چکیده مقاله:

Background: Small photon fields are increasingly used in modern radiotherapy especially in intensity modulated radiation therapy (IMRT) and stereotactic radiosurgery (SRS) treatments. Accurate beam profile and central axis depth doses measurements of such beams are complicated due to the electron disequilibrium. Hence the EBT2 (external beam therapy) Gafchromic film was used for dosimetry of small circular photon beams because of its high spatial resolution. The results of EBT2 film were compared with PinPoint measurements and Monte Carlo calculations. Materials and Methods: Four small field sizes (5, 10, 20 and 30 mm diameters) were produced by circular cones coupled to a Varian Clinac-2100 linear accelerator at 6MV photon beam energy. Experimental measurements were carried out using two dosimeters: The Gafchromic EBT2 film and the 0.015 cm3 PinPoint ionization chamber (PTW, type 31006). The EGSnrc/BEAMnrc code was used to calculate dosimetric parameters for the above small fields. Results: The results showed that the PDD values measured by EBT2 film had maximum local differences 4% and 3% with PinPoint measurements and MC calculations respectively. The measurements of penumbra (80-20%) at 5 cm depth in a water phantom and SSD=100 cm by EBT2 film were up to 0.8 and 0.4 mm smaller than PinPoint measurements and MC calculation respectively. Our results show good agreement between EBT2 film measurements and MC calculation for small field output factors but PinPoint measurements need correction factors. Conclusion: This study showed that the Gafchromic EBT2 film is suitable detector for small field dosimetry especially for penumbra and output factor measurements.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

small photon field dosimetry using ebt2 gafchromic film and monte carlo simulation

background: small photon fields are increasingly used in modern radiotherapy especially in intensity modulated radiation therapy (imrt) and stereotactic radiosurgery (srs) treatments. accurate beam profile and central axis depth doses measurements of such beams are complicated due to the electron disequilibrium. hence the ebt2 (external beam therapy) gafchromic film was used for dosimetry of sm...

متن کامل

Monte carlo simulation of varian clinac iX 10 MV photon beam for small field dosimetry

Background: The lack of lateral electronic disequilibrium (LED) becomes a main problem in small field. This factor affects the dose in target volume cannot predict correctly. In addition, utilization of high-energy linear accelerator (10 MV) can emit some unwanted particles (electron contamination). Therefore, the aim of this study was to characterize head linear accelerator (linac) Varian Clin...

متن کامل

On the Monte Carlo simulation of small-field micro-diamond detectors for megavoltage photon dosimetry.

Monte Carlo (MC) calculated detector-specific output correction factors for small photon beam dosimetry are commonly used in clinical practice. The technique, with a geometry description based on manufacturer blueprints, offers certain advantages over experimentally determined values but is not free of weaknesses. Independent MC calculations of output correction factors for a PTW-60019 micro-di...

متن کامل

A study of surface dosimetry for breast cancer radiotherapy treatments using Gafchromic EBT2 film

The present study quantified surface doses on several rectangular phantom setups and on curved surface phantoms for a 6 MV photon field using the Attix parallel-plate chamber and Gafchromic EBT2 film. For the rectangular phantom setups, the surface doses on a homogenous water equivalent phantom and a water equivalent phantom with 60 mm thick lung equivalent material were measured. The measureme...

متن کامل

SU-E-T-15: Small and Nonstandard Photon Field Dosimetry Characterization Using Monte Carlo Methods.

PURPOSE To quantify uncertainty reduction in small photon field dosimetry through characterization of ionization chambers and calibration conditions using detailed Monte Carlo methods benchmarked against NIST-traceable measurements. METHODS Phase space profiles were obtained using detailed EGSnrc Monte Carlo models for a Varian 6 MV photon linear accelerator, and a NIST-traceable cobalt-60 te...

متن کامل

Characteristic of Gafchromic Film for Small Photon Field

Small photon field used in the intensity modulated radiotherapy (IMRT) and stereotactic radio surgery (SRS). It is difficult to measure beam profile and output factor because of electron disequilibrium, partial shutting of source because of this challenges, we need the good detector. This article is collection of 5 or more article and summary of the results of EBT3 that compared with EBT2 film ...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ذخیره در منابع من قبلا به منابع من ذحیره شده

{@ msg_add @}


عنوان ژورنال

دوره 11  شماره None

صفحات  215- 224

تاریخ انتشار 2013-10

با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.

کلمات کلیدی

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023